RUN 46- started at 00:50
Horizontal beam position = 4.0 mm, vertical beam position = -5.0 mm.
Detectors are Left arm OUT = 36.64, right arm IN = 64.60 mm
Injection at ca. 6E7
Target density 6-7E9
-----
00:51
Merger OK, Tape server OK
ADC data item statistics looks as before. OK
FEE temperatures look OK
System checks OK
Leakage current stable
Attachments 1-2
01:19
Attachments 3-4 waveforms. DSSD0 stats fine
01:33
DSSD0 stats fine
Attachments 5-6 spectra and xy, but I am not sure at which time these are cause I am stupid and apparently the AIDAsort code was not running....
01:56
Restarted the AIDAsort code and noted that stats in DSSD0 were bad again. Another shake. This time no change to the beam on my side.
2:43
All checks fine. Attachments 7-10 (also spectra)
4:01
Notice stats on DSSD0 went bad again. Attachment 11. Gave detector a shake.
Now all system checks OK (Attachements 12-14)
4:19
Just noticed target is oscillating a bit, hitting 1E10. I moved it further in X and Y. Seems to not go below 6E9 in any case.
NOTE: target was on at x=-19899 y=500.
5:17
Merger OK, Tape server OK
ADC data item statistics looks as before. OK
FEE temperatures look OK
System checks OK
Leakage current stable
Attachments 15-18 (also waves)
5:53
Notice stats on DSSD0 went bad again. Gave detector a shake.
6:10
Stats ok in attachment 19, but
Noticed DSSD3 has started to get only a fate of 140 nstead of previous 5500is (similar to what DSSD2 has done earlier- see previous elog). Attachment 20 for the rates.
I have therefore made the Check ASIC control test from the ASIC control tab- attachment 21 with results.
DSSD3 back to normal operations - attachment 22 |