Message ID: 654
Entry time: Wed Jun 19 12:01:29 2024
|
Author: |
TD |
Subject: |
23:00-07:00 Wednesday 19 June |
|
|
23.18 DSSSD bias & leakage current OK - attachment 1
FEE64 temps OK - attachment 2
ADC data item stats attachment 3
aida01-aida04 c. 250k
per FEE64 1.8.W spectra - 20us FSR - attachments 4-5
large transients aida01-aida04
per FEE64 Rate spectra - attachment 6
Merger, TapeServer etc. - attachments 7-8
data file R101_15
DSSSD energy spectra (updating) - attachment 9
DSSSD bias & leakage current log - attachment 10
transients bottom & top left c. 02:00 18.6.24
cycle slow comparator 0x32 -> 0x64 -> Check ASIC Control -> 0x32
aida01 - aida04 rates return to normal
ADC data item stats - attachment 11
DSSSD energy spectra - attachment 12
23.40 cycle slow comparator 0x32 -> 0x64 -> Check ASIC Control -> 0x32
aida01 - aida04 rates return to normal
00.10 cycle slow comparator 0x32 -> 0x64 -> Check ASIC Control -> 0x32
aida01 - aida04 rates return to normal
01.43 ADC item stats OK - attachment 13
FEE64 temps OK - attachment 14
04.43 ADC data items stats
ADC data item stats - attachment 15
FEE64 temps OK - attachment 16
cycle slow comparator 0x32 -> 0x64 -> Check ASIC Control -> 0x32
aida01 - aida04 rates return to normal
ADC data item stats - attachment 17
per FEE64 Rate spectra - attachment 18
Merger, TapeServer etc. - attachment 19
data file R101_51
07.00 FEE64 temps OK - attachment 21
ADC data items OK - attachment 20
Merger, Tape Server OK - attachment 22
DSSSD bias & leakage current - attachment 23
DAQ stop, close data file
DSSSD bias OFF
FEE64 power OFF
Gzip'ing all uncompressed data files |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|