AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
CARME |
 |
|
Message ID: 667
Entry time: Mon Apr 7 08:59:41 2025
|
Author: |
TD |
Subject: |
Monday 7 April |
|
|
10.00 FEE64 power ON
CARME IE514 OFF
Confirm DAQ SETUP restores ASIC settings 2024Apr24-14.14.03 for *all* FEE64s
DSSSD bias & leakage current - attachment 1
leakage currents remain high cf. June 2024 - status of vacuum group controlled ion gauge unknown
12.10 FEE64 power ON
DSSSD bias & leakage current - attachment 2
FEE64 temps OK - attachment 3
All system wide checks OK
WR timestamps OK - attachment 4
per FEE64 Rate spectra - attachments 5-6
ADC data item stats - attachment 7
13.45 ML says vacuum group have switched off YR09DD1AS
DSSSD bias & leakage currents - attachment 8
DSSSD #2 and DSSSD #4 leakage currents return to 'normal' values cf. June 2024
DSSSD #1 and DSSSD #3 leakage currents reduce by c. 10-15% only and remain high cf. June 2024
17.47 DSSSD bias OFF
FEE6 Power OFF
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|