08:27
No run going.
Temperatures OK - attach
Attempting to bias channel 3 again. Slowly ramping up.
V = 60, I = 5.3
V = 70, I = 6.1
V = 80, I = 9.0 --> 6.2 (unstable). Decreasing bias.
V = 70, I = 6.0 (stable again)
Leaving at 70 V for a while to condition (?)
Resetting the gnuplot graph. Old caen_current.dat moved to G22-205-caen_current.dat and copied in /TapeServer/G22-205
10:10
Temperatures OK
Leakage current on channel 3 showed some issues, back to stable. See attached.
11:00
Temps OK
Leakage current stable.
Increased bias on channel 3 to 80 V. I = 6.2 uA stable. Leaving for a while.
12:45
Temps OK
Leakage current stable
Increased bias on channel 3 to 90 V. I = 6.3 stable. Leaving for a while.
13:30
Temps OK
Leakage current stable.
Increased bias on channel 3 to 100V. I = 6.3 stable. Hopefully it holds until the end of the beam time.
20:00
Leakage current on channel 3 had issues again and bias was reduced to 50 V. Remained stable. Turned to 60 V now. Still stable. See attachement.
FEEs on detector for -> MEC
Temps OK - attach
Checks as before - attach
Rates OK (four fees not counting as expected) - attach
Deuterium beam in the ring at 2 MeV/A approx 2E8 pps, significant H2+ contamination.
Deuterium target working, fluctuating around 5E12-1E13 atoms/cm2
Vacuum in YR09 - 3E-10 mbar
Beginning beam-target overlap measurements soon. |