216
|
Mon Feb 21 09:07:37 2022 |
RSS, JG | Alpha background run50 |
Beam OFF. Run started 20/2/22 18:42
System checck OK, except white rabbit all fail
Temperatures OK
Leakage current 3.57 uA.
potentiometer = 20.28 mm
|
215
|
Mon Feb 21 08:05:45 2022 |
RSS, JG | Alpha background run50 |
Beam OFF. Run started 20/2/22 18:42
System checck OK, except white rabbit all fail
Temperatures OK
Leakage current 3.57 uA.
potentiometer = 20.28 mm
|
214
|
Sun Feb 20 22:04:21 2022 |
JM, CB | Alpha Background run 50 |
Beam OFF. Run started 20/2/22 18:42
Fast comparator threshold HEC set to 0xff for all fees and asics to combat high energy noise bursts.
Electron cooler scalar plugged back into aida03 ~3000 per sec.
System checck OK, except white rabbit all fail
Temperatures OK
Leakage current 3.638 uA.
Statistics rates in noraml range, aida03 slightly higher as expected with scalar. |
213
|
Sat Feb 19 20:08:43 2022 |
JM, CB | Beam ON run 49 |
Run started 19/2/22 21:05. Beam ON
Temperatures OK
System wide checks OK - white rabbit fails on all.
Statistics are within normal range (10-20K goodevents)
Leakage current 3.6 uA
potentiometer = 20.28 mm
20/2/22
Leakage current ~3.675 uA. Looks stable now motors can not move (attachment 1)
potentiometer = 20.28 mm
Temperatures OK
System wide checks OK - white rabbit fails all
Statistic within normal range (see attachment 2 - ADC data items) Counter for aida 1 and aida 2 much higher due to noise bursts (elog 212)
XY total plot (attachment 3)
Low energy Ex histogram (attachment 4)
Low energy Ey histogram (attachment 5)
Tape Server data written (attachment 6)
Run end 20/2/22 15:08 |
Attachment 1: 2022-02-20_14-16-16-leakage-current.png
|
|
Attachment 2: 2022-02-20_14-18-07-statistic-ADCdata.png
|
|
Attachment 3: 2022-02-20_14-20-43-xyTotal.png
|
|
Attachment 4: 2022-02-20_14-24-19-LowEnergy-ExTotal.png
|
|
Attachment 5: 2022-02-20_14-28-02-LowEnergy-EyTotal.png
|
|
Attachment 6: 2022-02-20_14-25-45-Tape-server-data-written.png
|
|
212
|
Sat Feb 19 14:51:58 2022 |
JM, CB | Burst high energy noise |
Noticed bursts of high energy noise on aida01 and aida02 during run 47 and 48.
Good events rate jumps from 10-20k range to 250k. ADC data items rate jumps from 0-1 to 2k. Data rate being written increases to ~2500 kb/s from ~20 kb/s
Effect appears short lived ~30s at a time. On 18/2/22 the asic was check loaded which appeared to remove the increased rate but due to the time taken to asic check the burst had already finished.
Asic 2 on aida01 and asic 1 on aida02 are affected. See attached plots.
Fast comparator threshold HEC currently set to 0x2 for all
Zoomed in region of aida 1 asic 2 and the same region for asic 1 to see the difference are attached.
|
Attachment 1: 2022-02-19_15-50-50-aida1-4-stat-logscale.png
|
|
Attachment 2: 2022-02-19_15-47-13-aida1-2*H.png
|
|
Attachment 3: 2022-02-19_15-49-48-aida2-1*H.png
|
|
Attachment 4: 2022-02-20_14-01-40-aida1-2*H.png
|
|
Attachment 5: 2022-02-20_14-02-55-aida1-1*H.png
|
|
211
|
Fri Feb 18 20:31:33 2022 |
JM, CB | Alpha Background run 48 |
Alpha background started with beam ON - run47. Run started 16:49 18/2/22 (beam subsequently found to be OFF)
Beam energy 4 keV, beam lifetime ~ 100 ms
Noticed aida01 and aida02 were showing increased good event rates to ~250,000, effect seen on spectrum browser and data writing rate increased to ~2500 kb/sec from ~20 kb/sec. ASIC check on the affected fees brought rates back to normal range 10-20k good events. Multiple instances occurred in a row on run47 so fees were power cycled. Run ended 17:21
Run 48 started after power cycle
Run start 17:42, beam OFF ~18:30-19:00 for access to the cave to aid allignment of the gas jet target.
Rates remain stable until ~20:00. Aida01 and 02 show increased rates, asic check -> back to normal range.
Beam ON 21:43
All system checks OK. All rates OK. Leakage current ~3.62 uA.
14:00 19/2/22
Beam OFF during 18/2/22 due to an interlock controlled by main control room unknown to us. Beam has been OFF.
Leakage current plot attached. System checks OK, except white rabbit, all fail.
Motors were at 26mm potentiometer opposed to 20.2mm which was the set point (fully out 19mm). Motors dragged in as pneumatic was not fully out. Likely result of pneumatic out valve being closed. Pneumatic fully out, potentometer back to 20.2mm and leakage current drop from ~3.75 uA to 3.54uA.
Run ended 21:00 |
Attachment 1: 2022-02-19_14-45-25-leage-current.png
|
|
210
|
Fri Feb 18 13:03:02 2022 |
RSS, ML, CB | Alpha background run46 |
14:00
Check Leakage current: (see attachment 4)
Check NewMerger tab: (see attachment 2)
Check Fee temperatures: all green (see attachment 1)
Check statistics tab: ADC data items (see attachment 11)
ADC coincidence-info #8 (see attachment 10)
Good events (see attachment 9)
Check Spectrum Browser tab: (see attachment 6)
Check the following plots:
LowEnergyXYTotalDSSD0 (see attachment 7)
LowEnergyExTotal (see attachment 8)
LowEnergyEyTotal (see attachment 5)
Check system wide checks tab: White Rabbit error counter test result: Passed 0, Failed 5 (see attachment 3)
Data taking was stopped at 14:04. |
Attachment 1: temp-2022-02-18_13-59-30.png
|
|
Attachment 2: merger-2022-02-18_13-58-41.png
|
|
Attachment 3: system-2022-02-18_13-59-57.png
|
|
Attachment 4: leakage-2022-02-18_13-59-02.png
|
|
Attachment 5: ey-total-2022-02-18_13-57-46.png
|
|
Attachment 6: spectrum-2022-02-18_14-02-02.png
|
|
Attachment 7: xy-total-2022-02-18_14-02-23.png
|
|
Attachment 8: ex-energy-2022-02-18_13-56-32.png
|
|
Attachment 9: good-event-2022-02-18_14-00-53.png
|
|
Attachment 10: correlation-2022-02-18_14-00-27.png
|
|
Attachment 11: adc-data-item-2022-02-18_14-01-17.png
|
|
209
|
Fri Feb 18 11:07:52 2022 |
RSS, ML, CB | Alpha background run46 |
12:00
Check Leakage current: (see attachment 4)
Check NewMerger tab: (see attachment 2)
Check Fee temperatures: all green (see attachment 1)
Check statistics tab: ADC data items (see attachment 9)
ADC coincidence-info #8 (see attachment 8)
Good events (see attachment 7)
Check Spectrum Browser tab: (see attachment 10)
Check the following plots:
LowEnergyXYTotalDSSD0 (see attachment 11)
LowEnergyExTotal (see attachment 5)
LowEnergyEyTotal (see attachment 6)
Check system wide checks tab: White Rabbit error counter test result: Passed 0, Failed 5 (see attachment 3) |
Attachment 1: temp-2022-02-18_12-01-21.png
|
|
Attachment 2: merger-2022-02-18_12-00-10.png
|
|
Attachment 3: system-2022-02-18_12-01-52.png
|
|
Attachment 4: leakage-2022-02-18_12-00-43.png
|
|
Attachment 5: energy-ex-2022-02-18_12-07-07.png
|
|
Attachment 6: energy-ey-2022-02-18_12-05-29.png
|
|
Attachment 7: good-event-2022-02-18_12-02-49.png
|
|
Attachment 8: correlation-2022-02-18_12-03-19.png
|
|
Attachment 9: adc-data-item2022-02-18_12-02-25.png
|
|
Attachment 10: spectrum_browser-2022-02-18_12-03-55.png
|
|
Attachment 11: energy-xy-total-2022-02-18_12-14-29.png
|
|
208
|
Fri Feb 18 08:58:08 2022 |
RSS, ML, CB | Alpha background run46 |
10:00
Check Leakage current: (see attachment 2)
Check NewMerger tab: (see attachment 5)
Check Fee temperatures: all green (see attachment 1)
Check statistics tab: ADC data items (see attachment 3)
ADC coincidence-info #8 (see attachment 6)
Good events (see attachment 7)
Check Spectrum Browser tab: (see attachment 9)
Check the following plots:
LowEnergyXYTotalDSSD0 (see attachment 9)
LowEnergyExTotal (see attachment 10)
LowEnergyEyTotal (see attachment 11)
Check system wide checks tab: White Rabbit error counter test result: Passed 0, Failed 5 (see attachment 8) |
Attachment 1: temp-2022-02-18_09-50-32.png
|
|
Attachment 2: leakage-2022-02-18_09-44-34.png
|
|
Attachment 3: adc-data-2022-02-18_09-52-48.png
|
|
Attachment 4: energy_ex_2022-02-18_09-55-48.png
|
|
Attachment 5: new-merger-2022-02-18_09-44-10.png
|
|
Attachment 6: correlation-2022-02-18_09-51-48.png
|
|
Attachment 7: good-events-2022-02-18_09-52-17.png
|
|
Attachment 8: system-wide-2022-02-18_09-51-08.png
|
|
Attachment 9: spectrum-browser-2022-02-18_09-53-52.png
|
|
Attachment 10: energy-xy-total-2022-02-18_10-12-10.png
|
|
Attachment 11: energy-ey-2022-02-18_09-58-53.png
|
|
207
|
Thu Feb 17 18:19:16 2022 |
JM, CB, SD | Alpha Background run 46 |
Both GSI vacuum group and edinburgh ion gauges OFF
Fees power cycled. All system checks OK.
Writing to file, alpha background overnight.
Moved detector IN 2mm. Noise appears reduced with any distance moved in so overnight spectra to be taken while IN. Not moving further in as possible beam tomorrow.
22:44
Leakage current stable, slowly decreasing ~3.57 uA (attachment 1)
Statistics look OK
Temperatures OK
System wide check all pass - OK
Noise appears much reduced on spectrum Browser channels. (attachement 2). Compared to same plot on elog 197
XYtotal per pixel (attachment 3)
Ex histogram (attachment 4)
|
Attachment 1: 2022-02-17_22-41-45-leakageCurrent.png
|
|
Attachment 2: 2022-02-17_22-51-07-aida1-3*L.png
|
|
Attachment 3: 2022-02-17_22-54-06-XYTotal.png
|
|
Attachment 4: 2022-02-17_22-56-38-ExHist.png
|
|
206
|
Thu Feb 17 16:26:27 2022 |
JM, CB, SD | Gauges OFF, Pulser Movement tests |
Edinburgh group ion gauge turned OFF -> decrease in leakage current from 10.05 uA to 6.5 uA
GSI vacuum group ion gauge turned OFF -> decrease in leakage current from 6.5 uA to 3.63 uA
GSI large ion pump turned ON -> no effect on leakage current
Motors moved IN the same as elog 201. No significant increase in leakage current as moved further into the chamber (attachment 1). Pulser widths follow similar path as previous tests. Aida1 and 2 statrt with larger pulser widths which reduce as are moved into the chamber. Aida3,4 start lower than aida1,2 and remain at same level throughout. (attachment2)
RF ring signal removed from aida5. Data rate dropped back to ~20 kb/sec, Merger rate ~80,000 items/sec
Noise peak on aida1 appears reduced after gauges switched OFF with detector fully out (attachment 3) coompare to elog 201
Noise peak is removed once detector is moved in. (attachment 4)
Ey histogram (attachment 5) shows multiple peaks for the same pulser amplitude.
XY total pixels plot (attachment 6)
1.8 waveforms for aida 1,2,5 (attachment 7) |
Attachment 1: leakage-potentiometer2_1.png
|
|
Attachment 2: pulserWidth_potentiometer2.png
|
|
Attachment 3: 2022-02-17_16-54-27aida125-1.8*L-fullyout.png
|
|
Attachment 4: 2022-02-17_16-48-55-aida125-1.8*L-fullyin.png
|
|
Attachment 5: 2022-02-17_17-29-25-Eyhistogram-pulserON.png
|
|
Attachment 6: 2022-02-17_18-40-14-xyTotal-pulser.png
|
|
Attachment 7: 2022-02-17_22-59-53-aida125-1.8*W.png
|
|
205
|
Thu Feb 17 15:09:40 2022 |
JM,CB,SD | Alpha background run44 |
16:09
Check Leakage current: (see attachment 1)
Check NewMerger tab: (see attachment 3)
Check Fee temperatures: all green (see attachment 5)
Check statistics tab: ADC data items (see attachment 2)
ADC coincidence-info #8 (see attachment 6)
Good events (see attachment 4)
Check Spectrum Browser tab: (see attachment 10)
Check the following plots:
LowEnergyXYTotalDSSD0 (see attachment 9)
LowEnergyExTotal (see attachment 7)
LowEnergyEyTotal (see attachment 8)
Check system wide checks tab: White Rabbit error counter test result: Passed 0, Failed 5 (see attachment 11)
Run stopped for pulser tests 16:13 |
Attachment 1: 2022-02-17_16-04-11_lowEnergyExTotal.png
|
|
Attachment 2: 2022-02-17_16-05-03_lowenergyEytotal.png
|
|
Attachment 3: 2022-02-17_16-03-06_lowEnergyTotalDSSSD.png
|
|
Attachment 4: 2022-02-17_15-54-12_leakagecurrent.png
|
|
Attachment 5: 2022-02-17_15-53-25_Merger.png
|
|
Attachment 6: 2022-02-17_15-52-29_Temperature.png
|
|
Attachment 7: 2022-02-17_15-50-57_GoodEvents.png
|
|
Attachment 8: 2022-02-17_15-49-12._AIDA_ADC_dataitems.png
|
|
204
|
Thu Feb 17 13:12:19 2022 |
RSS, AG | Alpha background run43 |
14:00
Check Leakage current: (see attachment 1)
Check NewMerger tab: (see attachment 3)
Check Fee temperatures: all green (see attachment 5)
Check statistics tab: ADC data items (see attachment 2)
ADC coincidence-info #8 (see attachment 6)
Good events (see attachment 4)
Check Spectrum Browser tab: (see attachment 10)
Check the following plots:
LowEnergyXYTotalDSSD0 (see attachment 9)
LowEnergyExTotal (see attachment 7)
LowEnergyEyTotal (see attachment 8)
Check system wide checks tab: White Rabbit error counter test result: Passed 0, Failed 5 (see attachment 11) |
Attachment 1: leakage_2022-02-17_13-58-44.png
|
|
Attachment 2: data_item_2022-02-17_14-02-32.png
|
|
Attachment 3: new-merger-2022-02-17_13-57-51.png
|
|
Attachment 4: good_events_2022-02-17_14-01-53.png
|
|
Attachment 5: temperature_2022-02-17_13-59-30.png
|
|
Attachment 6: correlation_8_2022-02-17_14-01-13.png
|
|
Attachment 7: energy_ex_total_2022-02-17_14-06-46.png
|
|
Attachment 8: energy_ey_total_2022-02-17_14-08-06.png
|
|
Attachment 9: energy-xy-total-2022-02-17_14-11-42.png
|
|
Attachment 10: spectrum_browser_2022-02-17_14-03-27.png
|
|
Attachment 11: system_wide_checks_2022-02-17_14-00-31.png
|
|
203
|
Thu Feb 17 08:31:07 2022 |
RSS, AG | Alpha background run43 |
09:15
Check Leakage current: (see attachment 2)
Check NewMerger tab: (see attachment 5)
Check Fee temperatures: all green (see attachment 1)
Check statistics tab: ADC data items (see attachment 4)
ADC coincidence-info #8 (see attachment 6)
Good events (see attachment 7)
Check Spectrum Browser tab: (see attachment 10)
Check the following plots:
LowEnergyXYTotalDSSD0 (see attachment 3)
LowEnergyExTotal (see attachment 8)
LowEnergyEyTotal (see attachment 9)
Check system wide checks tab: White Rabbit error counter test result: Passed 0, Failed 5 (see attachment 11)
|
Attachment 1: temp-2022-02-17_09-16-02.png
|
|
Attachment 2: leakage-2022-02-17_09-14-47.png
|
|
Attachment 3: XY-Total-2022-02-17_09-30-30.png
|
|
Attachment 4: data-items-2022-02-17_09-24-58.png
|
|
Attachment 5: new-merger-2022-02-17_09-14-00.png
|
|
Attachment 6: correlation-2022-02-17_09-24-05.png
|
|
Attachment 7: good-events-2022-02-17_09-21-48.png
|
|
Attachment 8: lowEnergy-Ex-2022-02-17_09-26-36.png
|
|
Attachment 9: lowEnergy-Ey-2022-02-17_09-27-46.png
|
|
Attachment 10: spectrum-browser-2022-02-17_09-25-45.png
|
|
Attachment 11: IMG_6932.JPG
|
|
202
|
Wed Feb 16 21:41:11 2022 |
JM | Run 43 Alpha Background |
22:41
Check Leakage current: (see attachment 1) = 10.01 uA and falling
Check NewMerger tab: (see attachment 2)
Check Tape Server: (see attachment 3) Writing ~2500 kb/s
Check Fee temperatures: all green
Check statistics tab: ADC data items (see attachment 4) -aida3 and aida4 reading zero. Appears to be low noise as stat histogram shows counts as does Ey and Ex low energy histograms
ADC coincidence-info #8 (see attachment 5) aida1,2,5 correlation #8 signals counted. Do not see aida3 correlation 8 signals. was 70-90 previously (elog201)
Good events (see attachment 6) see 10-20k on aida1-4.
Low energy histograms Ex, Ey (attachemnts 7,8)
Noise appears lower across all fees. Aida1 channels for asic 1 and 3 (noisiest) checked against elog 197, same channels are hot but at lower rate to elog 197.
|
Attachment 1: 2022-02-16_22-45-51-Leakage-Current.png
|
|
Attachment 2: 2022-02-16_22-43-15-NewMerger.png
|
|
Attachment 3: 2022-02-16_22-43-50-TapeServer.png
|
|
Attachment 4: 2022-02-16_22-49-51-ADC-data-items.png
|
|
Attachment 5: 2022-02-16_22-51-40-AIDA-correlation8.png
|
|
Attachment 6: 2022-02-16_22-48-47-GoodEvents.png
|
|
Attachment 7: 2022-02-16_22-52-51-lowenergyEx.png
|
|
Attachment 8: 2022-02-16_22-57-35-lowenergyEy.png
|
|
201
|
Wed Feb 16 21:27:31 2022 |
JM | Cryring signals, Pulser tests, Waveforms |
On 15/2/22 it was noted that the leakage current increases when the motors are moved IN. This is contary to tests conducted on 12/2/22 (elog 192) where no increase was observed.
16/2/22: Servo motor and pneumatic used to move in by different differences with leakage current observed to increase linearly with distance moved in, measured using the potentiometer. (attachment 10)
After each movement the pulser width was measured for each of aida1,2,3,4. No change observed for aida3,4 with pulser width ~220 channels. Pulser width decreased for aida1,2 when moved IN. (attachment 11)
Noise on aida 2,3,4 appeared low throughout tests. Aida1 has a large noise peak when in fully out position. When moved fully in, the noise peak disappears (attachments 1,2)
Cryring signals for ring RF, electron cooler voltage and schottky added to the motors signal and fed into mac b to be read in daq as aida correlation info #8 signal.
cryring bnc 13 - ring rf - aida5 ~ 500,000 rate
cryring bnc 14 - e cooler voltage - aida3 ~ 80 rate
cryring bnc 15 - shottky -aida2 ~2000 rate
cryring bnc 16 motors - aida1 ~3 rate
correlation info #8 appears to be added to the good events rate (attachments 4,5) ADC items (attachment 3)
Waveforms 1.8*W attached for 20us, 200us, 2ms, 20ms |
Attachment 1: 2022-02-16_16-01-32aida1-2-1.8*L-fullyout.png
|
|
Attachment 2: 2022-02-16_15-58-02aida1-2-1.8*L-fullyin.png
|
|
Attachment 3: 2022-02-16_18-10-14-ADCdataItems.png
|
|
Attachment 4: 2022-02-16_18-13-03-aidaCorrelation.png
|
|
Attachment 5: 2022-02-16_18-10-53-goodEvents.png
|
|
Attachment 6: 2022-02-16_18-28-35aida125-1.8*W-20us.png
|
|
Attachment 7: 2022-02-16_18-27-30aida125-1.8*W-200us.png
|
|
Attachment 8: 2022-02-16_18-25-11aida125-1.8*W-2ms.png
|
|
Attachment 9: 2022-02-16_18-23-50aida125-1.8*W-20ms.png
|
|
Attachment 10: leakage-potentiometer.png
|
|
Attachment 11: pulserWidth_potentiometer.png
|
|
200
|
Wed Feb 16 15:27:53 2022 |
EM, JM | Run 41 Pulser tests |
16:30
NOTE: Pulser is ON!
Check Leakage current: (see attachment 1)
Check NewMerger tab: (see attachment 2)
Check Fee temperatures: all green (see attachment 3)
Check statistics tab: ADC data items (see attachment 4)
ADC coincidence-info #8 (see attachment 5)
Good events (see attachment 6)
Check Spectrum Browser tab: (see attachment 7)
Check the following plots: LowEnergyXYRateDSSD0 (see attachment 8)
LowEnergyXYTotalDSSD0 (see attachment 9)
LowEnergyExTotal (see attachment 10)
LowEnergyEyTotal (see attachment 11)
Check system wide checks tab: all okay.
All is good. |
Attachment 1: 2022-02-16_16-05-20_leakagecurrent.png
|
|
Attachment 2: 2022-02-16_16-08-00_newmerger.png
|
|
Attachment 3: 2022-02-16_16-09-05_feetemperatures.png
|
|
Attachment 4: 2022-02-16_16-11-10_aida_adc_data.png
|
|
Attachment 5: 2022-02-16_16-12-20_aida01_goodevents.png
|
|
Attachment 6: 2022-02-16_16-13-29_aida01_correlationinfo.png
|
|
Attachment 7: 2022-02-16_16-16-43_spectrumbrowser.png
|
|
Attachment 8: 2022-02-16_16-21-39_lowenergyXYrate.png
|
|
Attachment 9: 2022-02-16_16-23-56_lowenergyXYtotal.png
|
|
Attachment 10: 2022-02-16_16-25-16_lowenergyEytotal.png
|
|
Attachment 11: 2022-02-16_16-25-53_lowenergyExtotal.png
|
|
199
|
Wed Feb 16 12:43:05 2022 |
TP, RSS | Alpha background run40 |
13:30
Check Leakage current: (see attachment 1)
Check NewMerger tab: (see attachment 2)
Check Fee temperatures: all green (see attachment 3)
Check statistics tab: ADC data items (see attachment 4)
ADC coincidence-info #8 (see attachment 5)
Good events (see attachment 6)
Check Spectrum Browser tab: (see attachment 7)
Check the following plots: LowEnergyXYRateDSSD0 (see attachment 8)
LowEnergyXYTotalDSSD0 (see attachment 9)
LowEnergyExTotal (see attachment 10)
LowEnergyEyTotal (see attachment 11)
Check system wide checks tab: all okay.
All is good. |
Attachment 1: leakage-2022-02-16_13-35-38.png
|
|
Attachment 2: merger-state-2022-02-16_13-35-16.png
|
|
Attachment 3: temperatures-2022-02-16_13-36-05.png
|
|
Attachment 4: aida-adc-data-2022-02-16_13-38-13.png
|
|
Attachment 5: adc-good-events-2022-02-16_13-37-02.png
|
|
Attachment 6: aida-correlation-2022-02-16_13-37-39.png
|
|
Attachment 7: spectrum-browser-2022-02-16_13-38-56.png
|
|
Attachment 8: lowenergy-xy-rate-dssd0-2022-02-16_13-42-12.png
|
|
Attachment 9: lowenergy-xytotal-dssd0-2022-02-16_13-41-31.png
|
|
Attachment 10: lowenergy-ex-total-dsssd0-2022-02-16_13-39-43.png
|
|
Attachment 11: lowenergy-ey-total-dsssd0-2022-02-16_13-40-12.png
|
|
198
|
Wed Feb 16 08:27:27 2022 |
TP, RSS | Alpha background run40 |
09:15
Check Leakage current: (see attachment 1)
Check NewMerger tab: (see attachment 2)
Check Fee temperatures: all green (see attachment 3)
Check statistics tab: ADC data items (see attachment 4)
ADC coincidence-info #8 (see attachment 5)
Good events (see attachment 7)
Check Spectrum Browser tab: (see attachment 6)
Check the following plots: LowEnergyExTotal (see attachment 8)
LowEnergyEyTotal (see attachment 9)
LowEnergyXYRateDSSD0 (see attachment 10)
LowEnergyXYTotalDSSD0 (see attachment 11)
Check system wide checks tab: all okay.
All is good. |
Attachment 1: leakage-2022-02-16_09-15-33.png
|
|
Attachment 2: merger-tab_2022-02-16_09-14-40.png
|
|
Attachment 3: temperatures-2022-02-16_09-16-35.png
|
|
Attachment 4: stats-adc-items-2022-02-16_09-20-16.png
|
|
Attachment 5: aida-correlation-2022-02-16_09-21-16.png
|
|
Attachment 6: spectrum-browser-2022-02-16_09-22-06.png
|
|
Attachment 7: stats-good-event-2022-02-16_09-18-13.png
|
|
Attachment 8: low-enegry-ex-tot-2022-02-16_09-23-25.png
|
|
Attachment 9: low-energy-ey-tot-2022-02-16_09-23-52.png
|
|
Attachment 10: low-energy-xy-rate-dsssd0-2022-02-16_09-26-43.png
|
|
Attachment 11: low-energy-xy-total-dssd0-2022-02-16_09-25-54.png
|
|
197
|
Tue Feb 15 21:28:14 2022 |
JM | Noisy Channels |
All fees and asics checked to find noisy channels. Channels are listed along with the noise peak height for the low energy plot at 22:00. Run started ~17:45.
Attachment 1 shows 3*L for aida01 with x axis zoomed in on noise peak. Attachment 2 shows stat for aida01-04 in log scale.
Aida01
- 1.0L ~3000
- 1.2L ~3500
- 1.7L ~3000
- 1.8L ~2000
- 2.11L ~2000
- 2.12L ~400
- 3.7L ~1000
- 3.10 ~60000
- 3.11L ~2000
- 3.12L ~15000
- 4.12L ~400
- 4.14L ~750
Aida02
- 1.0L ~1000
- 1.3L ~750
- 1.7L ~600
- 1.15L ~600
- 4.15 ~125
Aida03
nothing noisy of note on aida04 |
Attachment 1: 2022-02-15_22-17-51-aida1-3*L.png
|
|
Attachment 2: 2022-02-15_22-23-12-aida1-4-stat-log.png
|
|