Entry  Mon Feb 21 17:26:19 2022, JM, CB, SHB, OH (remotely), Beam ON run 53 9x

21/2/22 16:00

Call from main control room, beam is ON, probably has been for several hours judging by rate on xy total plot.

Changes made to online monitoring code by OH, include setting to only 1 DSSD, correct offsets from the pulser walkthrough, offline analysis capabilities and X/Y position correction. 

Monitoring code restarted in CARME_SORT/build_no_output

17:31 Run 50 ended, run 51 started now beam is definitely in the ring. Tried to insert span zero schottky into aida2 correlation scalar, however rate was far too high.

18:03 Removed span zero schottky signal, started new run 52.

With beam ON we noticed some missing channels in the xy total plot. Aida 1 and 2 stat totals on spectrum browser show missing channels. Asic check on both to no effect. Fees power cycled.

19:57 Started run 53. Currently waiting for beam

21:20 Beam ON

1*L plots checked for aida 1, several channels showed no data (noise)

Slow comparator LEC threshold lowered to 0x32  to aida 1, checked 1*L plots and channels are now showing lots of noise. So those channels appear to be quiet using 0x64 threshold. (attachment 3)

Threshold increased back to 0x64

22:33

System checks OK -white rabbit fails on all. ~30 seperation

Temperatures OK

Statistics within range - new scalars added for aida02 and aida05

Data being written ~300 kb/sec

Leakage current ~3.6 uA

 

00:39

System checks checks OK -white rabbit fails on all. ~42 seperation

Temperatures OK

Statistics within range with new scalars

Data being written 256 kb/sec

Leakage current ~3.615 uA

XY counts dominated by hot channels - most blue pixels are ~10 counts

 

Entry  Mon Feb 21 12:10:47 2022, RSS, JG, Alpha background run50 220221_1506_aida03_1*W.png

Beam OFF. Run started 20/2/22 18:42

System checck OK, except white rabbit all fail

Temperatures OK

Leakage current 3.59 uA.

potentiometer = 20.28 mm

 

11:36 OH noticed a channel in aida03 was running hot. Did an asic load check and rate returned to normal.

          Online monitor also restarted.

 

15:06 Investigating the high rate in aida03 ASIC01 channel 11.
          Looking at the waveforms it seems channel 11 is consistently different.
          Attachment 1 shows a typical waveform for it

 

Entry  Mon Feb 21 10:59:48 2022, RSS, JG, Alpha background run50 

Beam OFF. Run started 20/2/22 18:42

System checck OK, except white rabbit all fail

Temperatures OK

Leakage current 3.59 uA.

potentiometer = 20.28 mm

 

11:36 OH noticed a channel in aida03 was running hot. Did an asic load check and rate returned to normal.

          Online monitor also restarted.

 

Entry  Mon Feb 21 09:07:37 2022, RSS, JG, Alpha background run50 

Beam OFF. Run started 20/2/22 18:42

System checck OK, except white rabbit all fail

Temperatures OK

Leakage current 3.57 uA.

potentiometer = 20.28 mm

 

Entry  Mon Feb 21 08:05:45 2022, RSS, JG, Alpha background run50 

Beam OFF. Run started 20/2/22 18:42

System checck OK, except white rabbit all fail

Temperatures OK

Leakage current 3.57 uA.

potentiometer = 20.28 mm

 

Entry  Sun Feb 20 22:04:21 2022, JM, CB, Alpha Background run 50  

Beam OFF. Run started 20/2/22 18:42

Fast comparator threshold HEC set to 0xff for all fees and asics to combat high energy noise bursts.

Electron cooler scalar plugged back into aida03 ~3000 per sec.

System checck OK, except white rabbit all fail

Temperatures OK

Leakage current 3.638 uA.

Statistics rates in noraml range, aida03 slightly higher as expected with scalar.

Entry  Sat Feb 19 20:08:43 2022, JM, CB, Beam ON run 49 6x

Run started 19/2/22 21:05. Beam ON

Temperatures OK

System wide checks OK - white rabbit fails on all.

Statistics are within normal range (10-20K goodevents)

Leakage current 3.6 uA

potentiometer = 20.28 mm

 

20/2/22

Leakage current ~3.675 uA. Looks stable now motors can not move (attachment 1)

potentiometer = 20.28 mm

Temperatures OK

System wide checks OK - white rabbit fails all

Statistic within normal range (see attachment 2 - ADC data items) Counter for aida 1 and aida 2 much higher due to noise bursts (elog 212)

 

XY total plot (attachment 3)

Low energy Ex histogram (attachment 4)

Low energy Ey histogram (attachment 5)

Tape Server data written (attachment 6)

 

Run end 20/2/22 15:08

Entry  Sat Feb 19 14:51:58 2022, JM, CB, Burst high energy noise 2022-02-19_15-50-50-aida1-4-stat-logscale.png2022-02-19_15-47-13-aida1-2*H.png2022-02-19_15-49-48-aida2-1*H.png2022-02-20_14-01-40-aida1-2*H.png2022-02-20_14-02-55-aida1-1*H.png

Noticed bursts of high energy noise on aida01 and aida02 during run 47 and 48.

Good events rate jumps from 10-20k range to 250k. ADC data items rate jumps from 0-1 to 2k. Data rate being written increases to ~2500 kb/s from ~20 kb/s

Effect appears short lived ~30s at a time. On 18/2/22 the asic was check loaded which appeared to remove the increased rate but due to the time taken to asic check the burst had already finished.

Asic 2 on aida01 and asic 1 on aida02 are affected. See attached plots.

Fast comparator threshold HEC currently set to 0x2 for all

Zoomed in region of aida 1 asic 2 and the same region for asic 1 to see the difference are attached.

 

Entry  Fri Feb 18 20:31:33 2022, JM, CB, Alpha Background run 48 2022-02-19_14-45-25-leage-current.png

Alpha background started with beam ON - run47. Run started 16:49 18/2/22   (beam subsequently found to be OFF)

Beam energy 4 keV, beam lifetime ~ 100 ms

Noticed aida01 and aida02 were showing increased good event rates to ~250,000, effect seen on spectrum browser and data writing rate increased to ~2500 kb/sec from ~20 kb/sec. ASIC check on the affected fees brought rates back to normal range 10-20k good events. Multiple instances occurred in a row on run47 so fees were power cycled. Run ended 17:21

Run 48 started after power cycle

Run start 17:42, beam OFF ~18:30-19:00 for access to the cave to aid allignment of the gas jet target.

Rates remain stable until ~20:00. Aida01 and 02 show increased rates, asic check -> back to normal range.

Beam ON 21:43

All system checks OK. All rates OK. Leakage current ~3.62 uA.

14:00 19/2/22

Beam OFF during 18/2/22 due to an interlock controlled by main control room unknown to us. Beam has been OFF. 

Leakage current plot attached. System checks OK, except white rabbit, all fail.

Motors were at 26mm potentiometer opposed to 20.2mm which was the set point (fully out 19mm). Motors dragged in as pneumatic was not fully out. Likely result of pneumatic out valve being closed. Pneumatic fully out, potentometer back to 20.2mm and leakage current drop from ~3.75 uA to 3.54uA.

Run ended 21:00

Entry  Fri Feb 18 13:03:02 2022, RSS, ML, CB, Alpha background run46 11x

14:00

 

Check Leakage current: (see attachment 4)

Check NewMerger tab: (see attachment 2)

Check Fee temperatures: all green (see attachment 1)

Check statistics tab: ADC data items (see attachment 11)

                                 ADC coincidence-info #8 (see attachment 10)

                                 Good events (see attachment 9)

 Check Spectrum Browser tab: (see attachment 6)                 

Check the following plots:

                                         LowEnergyXYTotalDSSD0 (see attachment 7)

                                         LowEnergyExTotal (see attachment 8)

                                         LowEnergyEyTotal (see attachment 5)

                                        

                                       

Check system wide checks tab: White Rabbit error counter test result: Passed 0, Failed 5 (see attachment 3)

Data taking was stopped at 14:04.

Entry  Fri Feb 18 11:07:52 2022, RSS, ML, CB, Alpha background run46 11x

12:00

 

Check Leakage current: (see attachment 4)

Check NewMerger tab: (see attachment 2)

Check Fee temperatures: all green (see attachment 1)

Check statistics tab: ADC data items (see attachment 9)

                                 ADC coincidence-info #8 (see attachment 8)

                                 Good events (see attachment 7)

 Check Spectrum Browser tab: (see attachment 10)                 

Check the following plots:

                                         LowEnergyXYTotalDSSD0 (see attachment 11)

                                         LowEnergyExTotal (see attachment 5)

                                         LowEnergyEyTotal (see attachment 6)

                                        

                                       

Check system wide checks tab: White Rabbit error counter test result: Passed 0, Failed 5 (see attachment 3)

Entry  Fri Feb 18 08:58:08 2022, RSS, ML, CB, Alpha background run46 11x

10:00

 

Check Leakage current: (see attachment 2)

Check NewMerger tab: (see attachment 5)

Check Fee temperatures: all green (see attachment 1)

Check statistics tab: ADC data items (see attachment 3)

                                 ADC coincidence-info #8 (see attachment 6)

                                 Good events (see attachment 7)

 Check Spectrum Browser tab: (see attachment 9)                 

Check the following plots:

                                         LowEnergyXYTotalDSSD0 (see attachment 9)

                                         LowEnergyExTotal (see attachment 10)

                                         LowEnergyEyTotal (see attachment 11)

                                        

                                       

Check system wide checks tab: White Rabbit error counter test result: Passed 0, Failed 5 (see attachment 8)

Entry  Thu Feb 17 18:19:16 2022, JM, CB, SD, Alpha Background run 46 2022-02-17_22-41-45-leakageCurrent.png2022-02-17_22-51-07-aida1-3*L.png2022-02-17_22-54-06-XYTotal.png2022-02-17_22-56-38-ExHist.png

Both GSI vacuum group and edinburgh ion gauges OFF

Fees power cycled. All system checks OK.

Writing to file, alpha background overnight.

Moved detector IN 2mm. Noise appears reduced with any distance moved in so overnight spectra to be taken while IN. Not moving further in as possible beam tomorrow.

22:44

Leakage current stable, slowly decreasing ~3.57 uA (attachment 1)

Statistics look OK

Temperatures OK

System wide check all pass - OK

Noise appears much reduced on spectrum Browser channels. (attachement 2). Compared to same plot on elog 197

XYtotal per pixel (attachment 3)

Ex histogram (attachment 4)

 

Entry  Thu Feb 17 16:26:27 2022, JM, CB, SD, Gauges OFF, Pulser Movement tests 7x

Edinburgh group ion gauge turned OFF -> decrease in leakage current from 10.05 uA to 6.5 uA

GSI vacuum group ion gauge turned OFF -> decrease in leakage current from 6.5 uA to 3.63 uA

GSI large ion pump turned ON -> no effect on leakage current

Motors moved IN the same as elog 201. No significant increase in leakage current as moved further into the chamber (attachment 1). Pulser widths follow similar path as previous tests. Aida1 and 2 statrt with larger pulser widths which reduce as are moved into the chamber. Aida3,4 start lower than aida1,2 and remain at same level throughout. (attachment2)

RF ring signal removed from aida5. Data rate dropped back to ~20 kb/sec, Merger rate ~80,000 items/sec

Noise peak on aida1 appears reduced after gauges switched OFF with detector fully out (attachment 3) coompare to elog 201

Noise peak is removed once detector is moved in. (attachment 4)

Ey histogram (attachment 5) shows multiple peaks for the same pulser amplitude. 

XY total pixels plot (attachment 6)

1.8 waveforms for aida 1,2,5 (attachment 7)

Entry  Thu Feb 17 15:09:40 2022, JM,CB,SD, Alpha background run44 8x

16:09

 

Check Leakage current: (see attachment 1)

Check NewMerger tab: (see attachment 3)

Check Fee temperatures: all green (see attachment 5)

Check statistics tab: ADC data items (see attachment 2)

                                 ADC coincidence-info #8 (see attachment 6)

                                 Good events (see attachment 4)

 Check Spectrum Browser tab: (see attachment 10)                 

Check the following plots:

                                         LowEnergyXYTotalDSSD0 (see attachment 9)

                                         LowEnergyExTotal (see attachment 7)

                                         LowEnergyEyTotal (see attachment 8)

                                        

                                       

Check system wide checks tab: White Rabbit error counter test result: Passed 0, Failed 5 (see attachment 11)

Run stopped for pulser tests 16:13

Entry  Thu Feb 17 13:12:19 2022, RSS, AG, Alpha background run43 11x

14:00

 

Check Leakage current: (see attachment 1)

Check NewMerger tab: (see attachment 3)

Check Fee temperatures: all green (see attachment 5)

Check statistics tab: ADC data items (see attachment 2)

                                 ADC coincidence-info #8 (see attachment 6)

                                 Good events (see attachment 4)

 Check Spectrum Browser tab: (see attachment 10)                 

Check the following plots:

                                         LowEnergyXYTotalDSSD0 (see attachment 9)

                                         LowEnergyExTotal (see attachment 7)

                                         LowEnergyEyTotal (see attachment 8)

                                        

                                       

Check system wide checks tab: White Rabbit error counter test result: Passed 0, Failed 5 (see attachment 11)

Entry  Thu Feb 17 08:31:07 2022, RSS, AG, Alpha background run43 11x

09:15

Check Leakage current: (see attachment 2)

Check NewMerger tab: (see attachment 5)

Check Fee temperatures: all green (see attachment 1)

Check statistics tab: ADC data items (see attachment 4)

                                 ADC coincidence-info #8 (see attachment 6)

                                 Good events (see attachment 7)

 Check Spectrum Browser tab: (see attachment 10)                 

Check the following plots:

                                         LowEnergyXYTotalDSSD0 (see attachment 3)

                                         LowEnergyExTotal (see attachment 8)

                                         LowEnergyEyTotal (see attachment 9)

                                        

                                       

Check system wide checks tab: White Rabbit error counter test result: Passed 0, Failed 5 (see attachment 11)

 

Entry  Wed Feb 16 21:41:11 2022, JM, Run 43 Alpha Background 8x

22:41

Check Leakage current: (see attachment 1) = 10.01 uA and falling

Check NewMerger tab: (see attachment 2)

Check Tape Server: (see attachment 3) Writing ~2500 kb/s

Check Fee temperatures: all green

Check statistics tab: ADC data items (see attachment 4)  -aida3 and aida4 reading zero. Appears to be low noise as stat histogram shows counts as does Ey and Ex low energy histograms

                                 ADC coincidence-info #8 (see attachment 5) aida1,2,5 correlation #8 signals counted. Do not see aida3 correlation 8 signals. was 70-90 previously (elog201)

                                 Good events (see attachment 6) see 10-20k on aida1-4.

Low energy histograms Ex, Ey (attachemnts 7,8)

Noise appears lower across all fees. Aida1 channels for asic 1 and 3 (noisiest) checked against elog 197, same channels are hot but at lower rate to elog 197.

 

 

 

 

 

Entry  Wed Feb 16 21:27:31 2022, JM, Cryring signals, Pulser tests, Waveforms  11x

On 15/2/22 it was noted that the leakage current increases when the motors are moved IN. This is contary to tests conducted on 12/2/22 (elog 192) where no increase was observed. 

16/2/22: Servo motor and pneumatic used to move in by different differences with leakage current observed to increase linearly with distance moved in, measured using the potentiometer. (attachment 10)

After each movement the pulser width was measured for each of aida1,2,3,4. No change observed for aida3,4 with pulser width ~220 channels. Pulser width decreased for aida1,2 when moved IN. (attachment 11)

Noise on aida 2,3,4 appeared low throughout tests. Aida1 has a large noise peak when in fully out position. When moved fully in, the noise peak disappears (attachments 1,2)

 

Cryring signals for ring RF, electron cooler voltage and schottky added to the motors signal and fed into mac b to be read in daq as aida correlation info #8 signal. 

cryring bnc 13  - ring rf - aida5 ~ 500,000 rate

cryring bnc 14 - e cooler voltage - aida3 ~ 80 rate

cryring bnc 15 - shottky -aida2 ~2000 rate

cryring bnc 16 motors - aida1 ~3 rate   

correlation info #8 appears to be added to the good events rate (attachments 4,5) ADC items (attachment 3)

Waveforms 1.8*W attached for 20us, 200us, 2ms, 20ms

Entry  Wed Feb 16 15:27:53 2022, EM, JM, Run 41 Pulser tests 11x

16:30

NOTE: Pulser is ON!

Check Leakage current: (see attachment 1)

Check NewMerger tab: (see attachment 2)

Check Fee temperatures: all green (see attachment 3)

Check statistics tab: ADC data items (see attachment 4)

                                 ADC coincidence-info #8 (see attachment 5)

                                 Good events (see attachment 6)

 Check Spectrum Browser tab: (see attachment 7)                 

Check the following plots: LowEnergyXYRateDSSD0 (see attachment 8)

                                         LowEnergyXYTotalDSSD0 (see attachment 9)

                                         LowEnergyExTotal (see attachment 10)

                                         LowEnergyEyTotal (see attachment 11)

 

Check system wide checks tab: all okay.

 

All is good.

ELOG V3.1.4-unknown