AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  CIRCE  ELOG logo
Message ID: 42     Entry time: Wed Nov 21 08:37:24 2018
Author: Thomas Chillery, Marialuisa Aliotta, Raffaele Buompane 
Subject: Upstream Collimator 

On 19/11/2018:

During the target ladder calibration (see elog entry #41) we also replaced the collimator (thickness ~24mm, diameter 3mm) located immediately upstream from the target ladder - that is between the upstream and central sections of the chamber. We mounted an aluminium collimator (thickness ~1mm, diameter 5mm). See attached picture 1.

Using the same telescopic sight as that used for the target ladder calibration, we aligned the 5mm collimator with the beam axis.

Attachment 1: IMG_3391_Modified.JPG  3.954 MB  Uploaded Tue Dec 4 09:56:43 2018  | Hide | Hide all
IMG_3391_Modified.JPG
ELOG V3.1.4-unknown