On 28/11/2018:
Mounted 12um mylar in front of far S2 - see attached pictures 1 and 2
In air: Far S2 biased to -130.0V, leak I = 1.61uA. Diode biased to -70.1V, leak I = 0.01uA -> 0.00uA. Rate meter ~ 4500Hz.
Detectors switched off
17:50 started Ecodry 1 and 2, started opening manual valve.
18:00 Manual valve fully opened, T1600 backing valve opened, Turbos switched on.
Once pressures ~ 1e-5mBar, far S2 bias = -130.0V, leak I = 1.59uA. Diode bias = -70.1V, leak I = 0.00uA.
18:08 Pressures (mBar):
- TPU1 = 3.0e-5
- TPU2 = 6.0e-5
- TPD1 = 5.2e-6
- TPD2 = 8.7e-6
- CHUP = Sensor Off
- CHDW = No Sensor
Run # 181128181421
- Run time: 1799s
- Pulser ON, as before
- Alpha source
- S2 mylar: 12um
- DT ~ 12%
Started overnight run:
Run # 181128184505
- Run time: 51706s
- Pulser ON, as before
- Alpha source
- S2 mylar: 12um
- DT ~ 8%
- See attached picture 3 for sample far S2 front strip spectrum
18:50 End of day check:
Far S2 bias = -130.0V, leak I = 1.60uA.
Diode bias = -70.1V, leak I = 0.00uA.
Pressures (mBar):
- TPU1 = 3.8e-6
- TPU2 = 1.2e-6
- TPD1 = 1.4e-6
- TPD2 = 1.6e-6
- CHUP = Sensor Off
- CHDW = No Sensor
On 29/11/2018: Alpha run stopped
9:05: Far S2 bias = -130.0V, leak I = 1.70uA. Diode bias = -70.1V, leak I = 0.00uA.
Detector biases and preamps off.
Pressures (mBar):
- TPU1 = 7.6e-7
- TPU2 = 3.5e-7
- TPD1 = 7.7e-7
- TPD2 = 7.3e-7
- CHUP = Sensor Off
- CHDW = No Sensor
Closed manual and T1600 backing valves. Stopped turbos. Stopped Ecodry 1 and 2.
Switched off Nim bin, RAL crate, and CAMAC crate
9:45 Opened venting valve |