Message ID: 8     Entry time: Wed Feb 22 14:06:47 2017
Author: HS 
Subject: NaI preparation 

Today we took datas of NaI detectors one by one for calibration. The gamma ray source was put on the NaI surfaces so that peaks could clealy appear where we expected.

We fitted them with gaussian and exponential, and integrate the peaks to calculate yields. The results were pretty good. 

However, at the test experiment last year, some NaI gains gradually shifted lower and peaks moved left during the beam time. So we need to let threshold level of CFD down NaI2 and NaI3 and NaI5, or change the fast amp gain and use downscaler, otherwise the part of peaks may go down below the threshold.

Probably NaI works have almost finished before setting up geometry I hope.

Attachment 1: NaICalibration_Cs137.pdf  491 kB  | Hide | Hide all
NaICalibration_Cs137.pdf
Attachment 2: NaICalibration.pdf  492 kB  Uploaded Wed Feb 22 15:14:34 2017  | Hide | Hide all
NaICalibration.pdf