AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DESPEC  ELOG logo
Message ID: 179     Entry time: Mon Mar 8 07:02:17 2021
Author: CA, LS 
Subject: Monday 8th March 08:00 - 17:00 
08:00 ASIC settings 2019Dec19-16.19.51
      DSSSD#1 slow comparator 0xa
      DSSSD#2 slow comparator 0xa
      DSSSD#3 slow comparator 0xd

      BNC PB-5 Pulser 
      Amplitude 1.0V
      Attenuation x1
      Frequency 2Hz
      tau_d 1ms
      - polarity
      Delay 250ns, tail pulse

      All system wide checks ok *except aida06 fails clock status 6*
      Statistics - attachment 1
      FEE Temperatures - attachment 2
      Bias and leakage currents ok - attachment 3

08:20 Merger ~4.5M data items/s
      TapeServer ~ 45MB/s

      Writing to file R6_1148

09:00 Beam off for optimisation - writing to file R6_1212

      Expected back at 09:30

09:30 Beam will be off for a bit longer - an hour or two

09:34 All system wide checks ok *except aida06 fails clock status 6*
      Statistics - attachment 4
      FEE Temperatures - attachment 5
      Bias and leakage currents ok - attachment 6

      Merger ~4.5M data items/s
      TapeServer ~ 45MB/s

10:10 (NH) Stopped DAQ to try the new merger from VP, new run R7, R8 for these tests
      Tape rate was 110 MB/s (!) and MBS relay seemed nothing... also it didn't stop when DAQ was stopped but took 10+ seconds 
to finish
      Something confusing, will look at data and report to VP
      Reverted to original merger and back... 45 MB/s
      R9

12.00 (LS)
      System wide checks okay except 'FEE64 module aida06 global clocks failed, 6'
      Statistics (attachment7)
      FEE temps (attatchment8)
      Rate spectra (attachment9), beam still down
      Leakage Currents written to sheets (attachment10)
      Merger ~45M items/s
      Tapeserver ~43MB/s

14.00 Still no beam but is expected to return soon
      System wide checks okay except 'FEE64 module aida06 global clocks failed, 6'
      Statistics (attachment11)
      FEE temps (attatchment12)
      Leakage Currents written to sheets (attachment13)
      Merger ~45M items/s
      Tapeserver ~44MB/s

14.40 Leakage currents still rising slowly compared to previous days, so will keep an eye on them (attachment14)

15.00 Beam is back (attachment15)
      Still writing to R9 (roughly around R9_386)

15.28 No beam (attachment16)

15.41 Beam back (attachment17)
      Roughly R9 (roughly around R9_439)

16.00 Analysis of R9_469, from FEE0 and FEE2 the hit rate for one side is around 48 HEC data items/second (attachment18)
 
16.20 System wide checks okay except 'FEE64 module aida06 global clocks failed, 6'
      Statistics (attachment19)
      Spectra Rate (attachment20)
      FEE temps (attatchment21)
      Leakage Currents written to sheets (attachment22), 
      still rising but looking on Grafana 7 day period should start to lower around 17.00
      Merger ~45M items/s
      Tapeserver ~43MB/s

16.55 Beam down (attachment23), back a couple minutes later

15.51 beam down (attachment24)
      
Attachment 1: Screenshot_from_2021-03-08_07-04-21.png  159 kB  | Hide | Hide all
Screenshot_from_2021-03-08_07-04-21.png
Attachment 2: Screenshot_from_2021-03-08_07-07-24.png  120 kB  | Hide | Hide all
Screenshot_from_2021-03-08_07-07-24.png
Attachment 3: Screenshot_from_2021-03-08_07-07-49.png  51 kB  | Hide | Hide all
Screenshot_from_2021-03-08_07-07-49.png
Attachment 4: Screenshot_from_2021-03-08_08-31-37.png  159 kB  Uploaded Mon Mar 8 08:35:51 2021  | Hide | Hide all
Screenshot_from_2021-03-08_08-31-37.png
Attachment 5: Screenshot_from_2021-03-08_08-32-58.png  120 kB  Uploaded Mon Mar 8 08:35:51 2021  | Hide | Hide all
Screenshot_from_2021-03-08_08-32-58.png
Attachment 6: Screenshot_from_2021-03-08_08-34-25.png  46 kB  Uploaded Mon Mar 8 08:35:51 2021  | Hide | Hide all
Screenshot_from_2021-03-08_08-34-25.png
Attachment 7: 1200_Statistics.PNG  126 kB  Uploaded Mon Mar 8 11:03:02 2021  | Hide | Hide all
1200_Statistics.PNG
Attachment 8: 1200_FEETemps.PNG  164 kB  Uploaded Mon Mar 8 11:03:05 2021  | Hide | Hide all
1200_FEETemps.PNG
Attachment 9: 1200_SpectraRate.PNG  222 kB  Uploaded Mon Mar 8 11:03:10 2021  | Hide | Hide all
1200_SpectraRate.PNG
Attachment 10: 1200_LeakageCurrents.PNG  31 kB  Uploaded Mon Mar 8 11:03:17 2021  | Hide | Hide all
1200_LeakageCurrents.PNG
Attachment 11: 1400_Statistics.PNG  127 kB  Uploaded Mon Mar 8 13:05:28 2021  | Hide | Hide all
1400_Statistics.PNG
Attachment 12: 1400_FEETemps.PNG  164 kB  Uploaded Mon Mar 8 13:05:34 2021  | Hide | Hide all
1400_FEETemps.PNG
Attachment 13: 1400_LeakageCurrents.PNG  31 kB  Uploaded Mon Mar 8 13:05:38 2021  | Hide | Hide all
1400_LeakageCurrents.PNG
Attachment 14: 1500_Grafana_LeakageCurrents.PNG  274 kB  Uploaded Mon Mar 8 14:09:09 2021  | Hide | Hide all
1500_Grafana_LeakageCurrents.PNG
Attachment 15: 1500_SpectraRate.PNG  225 kB  Uploaded Mon Mar 8 14:09:20 2021  | Hide | Hide all
1500_SpectraRate.PNG
Attachment 16: 1528_SpectraRate.PNG  222 kB  Uploaded Mon Mar 8 14:29:19 2021  | Hide | Hide all
1528_SpectraRate.PNG
Attachment 17: 1541_SpectraRate.PNG  225 kB  Uploaded Mon Mar 8 14:43:10 2021  | Hide | Hide all
1541_SpectraRate.PNG
Attachment 18: 1600_R9_469_hit_rates.PNG  117 kB  Uploaded Mon Mar 8 15:16:27 2021  | Hide | Hide all
1600_R9_469_hit_rates.PNG
Attachment 19: 1620_Statistics.PNG  126 kB  Uploaded Mon Mar 8 15:26:28 2021  | Hide | Hide all
1620_Statistics.PNG
Attachment 20: 1620_SpectraRate.PNG  224 kB  Uploaded Mon Mar 8 15:27:10 2021  | Hide | Hide all
1620_SpectraRate.PNG
Attachment 21: 1620_FEETemps.PNG  164 kB  Uploaded Mon Mar 8 15:27:18 2021  | Hide | Hide all
1620_FEETemps.PNG
Attachment 22: 1620_LeakageCurrents.PNG  31 kB  Uploaded Mon Mar 8 15:27:26 2021  | Hide | Hide all
1620_LeakageCurrents.PNG
Attachment 23: 1655_SpectraRate.PNG  222 kB  Uploaded Mon Mar 8 15:56:49 2021  | Hide | Hide all
1655_SpectraRate.PNG
Attachment 24: 1749_SpectraRate.PNG  221 kB  Uploaded Mon Mar 8 16:51:47 2021  | Hide | Hide all
1749_SpectraRate.PNG
ELOG V3.1.4-unknown