Message ID: 18
Entry time: Thu Jan 24 10:21:26 2019
|
Author: |
CA, TD, NH, VP |
Subject: |
January 24th 2019 |
|
|
10.30 AIDA setup and DAQ start complete
writing to file TapeData/NULL/R1
FEE64 Temperatures and Good event statistics ok (attachments 1 and 2)
11.00 results of analyser - shows significant number of timewarps in aida07, and a few in aida01-04
11.24 compared timestamps between mbs and AIDA, agree to 8 digits (attachment 3)
12.49 DAQ and Merger stopped and restarted
Raised slow comparator threshold from 0xa to 0x64
Raised fast comparator threshold (LEC/MEC) from 0xf to 0xff
12.51 Good event statistics: aida08 running at a normal rate
pulser peak (1.8.L) no longer has noise shoulder (attachment 4)
writing to file TapeData/NULL/R2
13.58 stopped writing to file. Pulser frequency increased to 1 kHz
writing to file TapeData/NULL/R3
15.00 file R4 - DAQ startup
17.58 trimmed and re-crimped ground wire on aida08
lowered pulser freq back to 50Hz
reduced slow comparator threshold to 0xa, fast comparator threshold (LEC) to 0xf
good event stats ok, aida08 no longer running hot (attachment 6)
1.8.L and 1.8.W spectra of aida08 (attachment 5)
pulser peak width now at 55.71
FEE64 temperatures ok (attachment 7) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|