21.04 DAQ continues OK - file R20_379
ASIC settings 2019Dec19-16.19.51
DSSSD#1 slow comparator 0xa
DSSSD#2 slow comparator 0xa
DSSSD#3 slow comparator 0xd
BNC PB-5 Pulser
Amplitude1.0V
Attenuation x1
Frequency 2Hz
tau_d 1ms
- polarity
Delay 250ns, tail pulse
System wide checks all OK *except*
ADC Calibration
FEE64 module aida07 failed
FEE64 module aida10 failed
Calibration test result: Passed 10, Failed 2
If any modules fail calibration , check the clock status and open the FADC Align and Control browser page to rerun calibration for that module
Check FPGA Timestamp Errors
Base Current Difference
aida12 fault 0x0 : 0x1 : 1
FPGA Timestamp error counter test result: Passed 11, Failed 1
If any of these counts are reported as in error
The ASIC readout system has detected a timeslip.
That is the timestamp read from the time FIFO is not younger than the last
FEE64 Temperatures OK - attachment 1
Good event statistics OK - attachment 2
Detector bias & leakage currents OK - attachment 3
Merger OK - 4.2M data items/s
TapeServer OK - 45 Mb/s
21.20 no merger server error/warning messages since 16.30UTC
21.35 p+n junction strip HEC & LEC spectra, Rate spectra common y-scale 0-30000 & 0-10
all spectra zero'd
22.05 no beam for 2-3h whilst accelerator generator is fixed |