16:25
Informed at shift handover that there has been no beam.
Statistics - See attachment 1
Temperatures - See attachment 2
Bias & Leakage currents - See attachment 3
System Wide checks - All passed, EXCEPT:
ADC Calibration
FEE64 module aida01 failed
FEE64 module aida02 failed
FEE64 module aida03 failed
FEE64 module aida04 failed
FEE64 module aida05 failed
FEE64 module aida06 failed
FEE64 module aida07 failed
FEE64 module aida08 failed
FEE64 module aida09 failed
FEE64 module aida10 failed
FEE64 module aida11 failed
FEE64 module aida12 failed
FEE64 module aida13 failed
FEE64 module aida14 failed
FEE64 module aida15 failed
FEE64 module aida16 failed
Calibration test result: Passed 0, Failed 16
If any modules fail calibration , check the clock status and open the FADC Align and Control browser page to rerun calibration for that module
White Rabbit decoder status:
Base Current Difference
aida05 fault 0x4da : 0x4dd : 3
White Rabbit error counter test result: Passed 15, Failed 1
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
18:00
Usecb scalars now showing DSSD implant rates.
Statistics - See attachment 4
Temperatures - See attachment 5
Bias & Leakage currents - See attachment 6
System Wide checks - All passed, EXCEPT:
ADC Calibration
FEE64 module aida01 failed
FEE64 module aida02 failed
FEE64 module aida03 failed
FEE64 module aida04 failed
FEE64 module aida05 failed
FEE64 module aida06 failed
FEE64 module aida07 failed
FEE64 module aida08 failed
FEE64 module aida09 failed
FEE64 module aida10 failed
FEE64 module aida11 failed
FEE64 module aida12 failed
FEE64 module aida13 failed
FEE64 module aida14 failed
FEE64 module aida15 failed
FEE64 module aida16 failed
Calibration test result: Passed 0, Failed 16
If any modules fail calibration , check the clock status and open the FADC Align and Control browser page to rerun calibration for that module
White rabbit decoder status
Base Current Difference
aida05 fault 0x4da : 0x4e1 : 7
White Rabbit error counter test result: Passed 15, Failed 1
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WRx
19:45
Usecb scalars have continued to show DSSD implant rates since the last recorded elog check.
Statistics - See attachment 7
Temperatures - See attachment 8
Bias & Leakage currents - See attachment 9
System Wide checks - All passed, EXCEPT:
ADC Calibration:
FEE64 module aida01 failed
FEE64 module aida02 failed
FEE64 module aida03 failed
FEE64 module aida04 failed
FEE64 module aida05 failed
FEE64 module aida06 failed
FEE64 module aida07 failed
FEE64 module aida08 failed
FEE64 module aida09 failed
FEE64 module aida10 failed
FEE64 module aida11 failed
FEE64 module aida12 failed
FEE64 module aida13 failed
FEE64 module aida14 failed
FEE64 module aida15 failed
FEE64 module aida16 failed
Calibration test result: Passed 0, Failed 16
If any modules fail calibration , check the clock status and open the FADC Align and Control browser page to rerun calibration for that module
White Rabbit decoder status:
Base Current Difference
aida05 fault 0x4da : 0x4e1 : 7
White Rabbit error counter test result: Passed 15, Failed 1
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
19:50 - Just as completing checks, notced usesb rates not showing DSSD implants again for several minutes
19:55 - Beam now back again - DSSD implants showing
|