AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DESPEC  ELOG logo
Message ID: 320     Entry time: Mon May 17 15:14:35 2021
Author: LJW 
Subject: Monday 17th May 2021 16:00-20:00 

16:25

Informed at shift handover that there has been no beam.

Statistics  - See attachment 1

Temperatures - See attachment 2

Bias & Leakage currents - See attachment 3

System Wide checks - All passed, EXCEPT:

ADC Calibration

   
FEE64 module aida01 failed
FEE64 module aida02 failed
FEE64 module aida03 failed
FEE64 module aida04 failed
FEE64 module aida05 failed
FEE64 module aida06 failed
FEE64 module aida07 failed
FEE64 module aida08 failed
FEE64 module aida09 failed
FEE64 module aida10 failed
FEE64 module aida11 failed
FEE64 module aida12 failed
FEE64 module aida13 failed
FEE64 module aida14 failed
FEE64 module aida15 failed
FEE64 module aida16 failed
Calibration test result: Passed 0, Failed 16

If any modules fail calibration , check the clock status and open the FADC Align and Control browser page to rerun calibration for that module

 

White Rabbit decoder status:

   
         Base         Current     Difference
aida05 fault      0x4da :      0x4dd :      3  
White Rabbit error counter test result: Passed 15, Failed 1

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

 

18:00

Usecb scalars now showing DSSD implant rates.

Statistics  - See attachment 4

Temperatures - See attachment 5

Bias & Leakage currents - See attachment 6

System Wide checks - All passed, EXCEPT:

ADC Calibration

   
FEE64 module aida01 failed
FEE64 module aida02 failed
FEE64 module aida03 failed
FEE64 module aida04 failed
FEE64 module aida05 failed
FEE64 module aida06 failed
FEE64 module aida07 failed
FEE64 module aida08 failed
FEE64 module aida09 failed
FEE64 module aida10 failed
FEE64 module aida11 failed
FEE64 module aida12 failed
FEE64 module aida13 failed
FEE64 module aida14 failed
FEE64 module aida15 failed
FEE64 module aida16 failed
Calibration test result: Passed 0, Failed 16

If any modules fail calibration , check the clock status and open the FADC Align and Control browser page to rerun calibration for that module

White rabbit decoder status

   
         Base         Current     Difference
aida05 fault      0x4da :      0x4e1 :      7  
White Rabbit error counter test result: Passed 15, Failed 1

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WRx

 

19:45

Usecb scalars have continued to show DSSD implant rates since the last recorded elog check.

Statistics  - See attachment 7

Temperatures - See attachment 8

Bias & Leakage currents - See attachment 9

System Wide checks - All passed, EXCEPT:

ADC Calibration:

   
FEE64 module aida01 failed
FEE64 module aida02 failed
FEE64 module aida03 failed
FEE64 module aida04 failed
FEE64 module aida05 failed
FEE64 module aida06 failed
FEE64 module aida07 failed
FEE64 module aida08 failed
FEE64 module aida09 failed
FEE64 module aida10 failed
FEE64 module aida11 failed
FEE64 module aida12 failed
FEE64 module aida13 failed
FEE64 module aida14 failed
FEE64 module aida15 failed
FEE64 module aida16 failed
Calibration test result: Passed 0, Failed 16

If any modules fail calibration , check the clock status and open the FADC Align and Control browser page to rerun calibration for that module


White Rabbit decoder status:

     Base         Current     Difference
aida05 fault      0x4da :      0x4e1 :      7  
White Rabbit error counter test result: Passed 15, Failed 1

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

 

19:50 - Just as completing checks, notced usesb rates not showing DSSD implants again for several minutes

19:55 - Beam now back again - DSSD implants showing

 

 

Attachment 1: 170521_1.png  74 kB  Uploaded Mon May 17 16:28:35 2021  | Hide | Hide all
170521_1.png
Attachment 2: 170521_2.png  109 kB  Uploaded Mon May 17 16:30:02 2021  | Hide | Hide all
170521_2.png
Attachment 3: 170521_3.png  9 kB  Uploaded Mon May 17 16:32:07 2021  | Hide | Hide all
170521_3.png
Attachment 4: 170521_4.png  74 kB  Uploaded Mon May 17 18:01:01 2021  | Hide | Hide all
170521_4.png
Attachment 5: 170521_5.png  109 kB  Uploaded Mon May 17 18:09:11 2021  | Hide | Hide all
170521_5.png
Attachment 6: 170521_6.png  9 kB  Uploaded Mon May 17 18:09:42 2021  | Hide | Hide all
170521_6.png
Attachment 7: 170521_7.png  74 kB  Uploaded Mon May 17 19:42:54 2021  | Hide | Hide all
170521_7.png
Attachment 8: 170521_8.png  110 kB  Uploaded Mon May 17 19:44:20 2021  | Hide | Hide all
170521_8.png
Attachment 9: 170521_9.png  9 kB  Uploaded Mon May 17 19:45:05 2021  | Hide | Hide all
170521_9.png
ELOG V3.1.4-unknown