Message ID: 427
Entry time: Tue May 10 07:41:28 2022
|
Author: |
NH TD |
Subject: |
Tuesday 10 May |
|
|
8:15am S4 CONTROLLED ACCESS to remove beam plug and check bPlas/AIDA response to beam
Reseat aida05 ribbon cable on adapter board while in...
No difference to rates, aida05 still shows hot asic01
aida14 seems seated OK, still asic02 basically empty
Noise level still high, threshold set to 0x20 (320 keV) which has decent rate
Start forwarding to MBS with 0x20 thresholds
10.12 slow comparator 0x20 -> 0xa
Attachments 1-5
Attachment 6 ADC data items
c. 150-180k per FEE64, c. 21Mb to disk
Attachment 7 FEE64 temps OK
Attachment 8 DSSSD bias & leakage currents OK
13.30 system checks
N.B. aida05 ASIC temp low and pause/resume data rate high
Attachment 9 grafana monitor of DSSSD bias & leakage current & temperature OK
Attachments 10-15 spectra
Attacahments 16-19 system wide checks OK *except* WR error counter
Attachment 20 ADC data items
c. 150-180k per FEE64, c. 21Mb to disk
Attachment 21 FEE64 temps OK
Attachment 22 DSSSD bias & leakage currents OK
Attachments 23-29 newMerger stats
20:37 - Begin implantation test, raise threshold to 0x64 just to ensure system is stable
22:26 - No implantation test although we had some brief beam through AIDA.
- FRS team having difficulty aligning beam still, will work overnight and try to implant tomorrow
Will keep DAQ on 0x64 to eliminate deadtime issues for implantation |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|