AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
DESPEC |
 |
|
Message ID: 430
Entry time: Thu May 12 00:37:52 2022
|
Author: |
OH, AM |
Subject: |
Thursday 12 May |
|
|
01:37 Over the past hour the degrader settings have been checked by impinging 209Pb onto the detectors.
That is now complete.
Will find optimum thresholds for n+n side and then leave DAQ running with n+n at optimum and p+n at 0xc
R16 slow comp 0x19 (250keV) 2 and 4 low deadtime 6 and 8 high deadtime. - attachment1
Will lower 2 and 4 to 200 and raise 6 and 8 to 300
R17 2 good, 4 shakey 6 and 8 poor - attachment 2
R18 2 and 4 at 220 and 6 and 7 at 350
2 and 8 low deadtime, 4 and 6 could be better. - attachment 3
Will run with p+n at 0xc 2+4 at 0x16 and 6 and 8 at 0x23
Statistics at this rate - attachment 4
02:18 Statistics - attachment 7
Temperatures - attachment 6
Bias & leakage current - attachment 5
Clock Status - OK
White Rabbit - OK
The other 3 checks currently aren't working
04:11 Statistics - attachment 10
Temperatures - attachment 9
Bias & leakage current - attachment 8
Clock Status - OK
White Rabbit - OK
06:05 Statistics - attachment 13
Temperatures - attachment 12
Bias & leakage current - attachment 11
Clock Status - OK
White Rabbit - OK |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|