AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DESPEC  ELOG logo
Message ID: 541     Entry time: Tue Mar 19 10:17:30 2024
Author: NH 
Subject: Dry Run 2024 - 19th March 24 
AIDA has 8 FEEs and 1 DSSSD 

Aida08 (HDMI#12) had no WR again, I moved it to a different MACB and now it gets WR
The MACB (currently with jsut HDMI#10) seems issues, check/replace the upstream HDMI and thent eh MACB (after dry run!)
 
DSSSD#1 biased to -120 V, leakage current 5.6 uA (fig 1)
 (also on Grafana)

Temps OK fig2

System Checks (fig3-5)
 Clocks OK
 aida07 fails calibration, others OK
 WR OK (aida04 0x10, seems OK)
 FPGA OK

Did "Synchronise ASIC clocks" to align ASIC clocks

(Notes for scalers: SC41L HDMI 5, SCI41R HDMI 9)

18:12 FRS is taking beam, AIDA is powered off and unbiased