Took over the shift from Marc.
10.51 Added screenshots to illustrate erroneous data from Grafana (AIDA) - attachments 18-19
Compare to the CAEN HV control terminal (attachment 15) which shows the expected values. Have email'd Nic & Helena. TD
10.55 Noticed that the CAEN HV control terminal is showing the corerct data but on the wrong line of the display (attachment 15). Enter commands q (quit) and d (display) to restart display - the correct data now appears on the correct line (attachment 20).
Grafana (AIDA) is now reporting data correctly too (attachment 21).
Summary - CAEN HV terminal display issue ( c. 10.05 - 10.55 )
Note the correct data/values continued to be shown. Leakage current shown correctly by both Grafana and CAEN HV control throughout.
Conclude this was a display issue - not a change in the operating parameters of the CAEN N1419ET.. Data should be OK.
14.45 analysis data file R21_500 - attachment 36
max deadtime 22% (aida11), 16% (aida04), 7% (aida02), 6% (aida08), 6% (aida06) , all others < 1%
no timewarps
HEC data item rate 2.7kHz
|