Message ID: 650
Entry time: Tue Jun 11 10:03:48 2024
|
Author: |
TD, NH |
Subject: |
08:00-16:00 Tuesday 10 June |
|
|
10.57 DSSSD bias & leakage current OK - attachment 1
FEE64 temps OK - attachment 2
ADC, MBS correlation scaler data item stats - attachments 3-4
ADC 9x < 20k, max 243k aida08
ASIC settings aida01 & aida02 - attachments 5-6
ALL system wide checks OK *except* attachments 7-8
WR timestamps OK - attachment 9
BNC PB-5 settings - attachment 10
per FEE64 Rate spectra - attachment 11
per p+n FEE64 1.8.L spectra - attachment 12
aida09 pulser peak width 57 ch FWHM
per FEE64 1.8.H spectra - attachments 13-14
per FEE64 1.8.W spectra - 20us FSR - attachments 15-16
no p+n FEE64 wavefroms - reason unknown
Merger, TapeServer etc - attachments 17-18
11.25 Lost control of aida-gsi window system
All command terminals lost, could not start new command terminals
aida-gsi restart
FEE64 power cycle
DSSSD bias cycle
Note file manager, calculator apps now working as expected
12.11 DAQ starts
data file S181/R6
ASIC settings 2024Apr19-15.22.49
BNC PB-5 Pulser
Amplitude 10.0V
Attenuation x10
Fall time 1ms
Frequency 2Hz
Polarity + (connected to test + signal daisy chain to all p+n FEE64s)
Sync ASIC clocks OK
All system wide checks OK *except* aida02 FPGA/WR decoder - attachments 19-20
per FEE64 1.8.W spectra - 20us FSR - attachments 21-22
12.29 analysis data file S181/R6_9 - attachment 23
max dead time c. 2% (aida08)
15:34 For some reason the AIDA MBS wasn't connecting to the time sorter (or a check stream) [NH]
Try to reset the AIDA MBS
Due to confusion the AIDA Merger was rebooted too, while the DAQ was running, this broke all the FEEs -> Powecycled
Synced ASIC clocks, fixed all waveforms
Masked all discriminators
All looks ok -> Sending back to MBS and back in DESPEC DAQ |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|