AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
DESPEC |
 |
|
Message ID: 664
Entry time: Sat Jun 15 00:28:01 2024
|
Author: |
TD |
Subject: |
Saturday 15 June |
|
|
Beam lost due to septum failure c. 18.15 Friday 14 June
Cooling of recirculating water failed c. 19.00 - water temperature increased from the normal 17-18deg C to 25-26 deg C
FEE64 temperatures increased and FEE64s powered off and on-call engineer requested.
Cooling of recirculating water restored and AIDA restarted
NH started background run 23.37 - data file S181/R9
01.21 DSSSD bias & leakage current OK - attachment 1
FEE64 temperatures OK - attachment 2
ADC data item stats OK - attachment 3
8x < 20k, max 238k aida08
per FEE64 Rate spectra - attachment 4
2x Check ASIC load
Merger, TapeServer etc - attachments 5-6
analysis data file R9_27 - attachment 7
max deadtime 1.9%, all other FEE64s < 1%
ADC data rate 951k, HEC data rate 1.6kHz
04.28 DSSSD bias & leakage current OK - attachment 8
FEE64 temperatures OK - attachment 9
ADC data item stats OK - attachment 10
8x < 20k, max 261k aida08
per FEE64 Rate spectra - attachment 11
08.30 DSSSD bias & leakage current OK - attachments 12-13
FEE64 temperatures OK - attachment 14
ADC data item stats OK - attachment 15
all rates increased - gamma source in use?
per FEE64 Rate spectra - attachments 16-17
note aida08 asic #3 hot HEC channel
per FEE64 1.8.L spectra - attachments 18-20
aida09 pulser peak width 59 ch FWHM
decay alphas aida05, aida12, aida02, aida04
per FEE64 1.8.W spectra - 20us FSR - attachments 21-23
Merger, TapeServer etc - attachments 25-25
analysis data file R9_152 - attachment 26
max deadtime 2.4%, all other FEE64s < 1%
ADC data rate 1.215M, HEC data rate 2.1kHz
12.22 DSSSD bias & leakage current OK - attachment 27
FEE64 temperatures OK - attachment 28
ADC data item stats OK - attachment 29
per FEE64 Rate spectra - attachment 30
15.27 DSSSD bias & leakage current OK - attachment 31
FEE64 temperatures OK - attachment 32
ADC data item stats OK - attachment 33
all rates increased - 370kBq 152Eu in use
per FEE64 Rate spectra - attachments 34
per FEE64 1.8.L spectra - attachments 35-36
aida09 pulser peak width 59 ch FWHM
decay alphas aida05, aida12, aida02, aida04
analysis data file R9_275 - attachment 37
max deadtime 4.4%, all other FEE64s < 2%
ADC data rate 1.520M, HEC data rate 2.3kHz
18.52 DSSSD bias & leakage current OK - attachment 38
FEE64 temperatures OK - attachment 39
ADC data item stats OK - attachment 40
per FEE64 Rate spectra - attachment 41
22.50 DSSSD bias & leakage current OK - attachment 42
FEE64 temperatures OK - attachment 43
ADC data item stats OK - attachment 44
per FEE64 Rate spectra - attachment 45 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|