AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DESPEC  ELOG logo
Entry  Mon Dec 16 13:07:04 2024, JB, MP, CC, AIDA timing test 16x
    Reply  Tue Dec 17 12:45:36 2024, JB, MP, CC, AIDA timing test IMG_2536.jpeg20241217_141148.jpg20241217_141206.jpg
    Reply  Wed Jan 8 16:47:05 2025, JB, MP, CC, AIDA timing test Screenshot_from_2025-01-08_17-51-51.png
       Reply  Thu Jan 9 12:46:43 2025, JB, MP, CC, AIDA timing test AIDAfasttime_bPlastWRdt_AIDA10.pngAIDACalAdcFasttime.pngAIDAfasttime_bPlastWRdt_AIDA10_nopulser.pngCentrepeak_gauss_fit.png
Message ID: 683     Entry time: Wed Jan 8 16:47:05 2025     In reply to: 677     Reply to this: 684
Author: JB, MP, CC 
Subject: AIDA timing test 

 

Quote:

https://elog.gsi.de/despec/Implantation+Stack/9?suppress=1 - Day 1 ELOG

Day2:

10:14 we set up the detector with the pulser in BB7 and started biasing the detector and setting up the DAQ. Water flow and temperature check, OK. 677/1 677/2 677/3

FEE temps OK. Screenshots included for LOCAL controls for aida07, Discriminator for aida07, ASIC control for aida01 ASIC # 1,2,3,4 and the pulser setting - 1 V @ 10 Hz rep. 677/4 677/5 677/6 677/7 677/8 677/9

ASIC thresholds for the fast comparator LEC/MEC set to 0x20 for all the ASICs in aida10 - connected to BB7. Act on all ASICs did not work with this so each threshold was set to 0x20 on each ASIC by hand.

The pulser is now connected to BB7 test + and via a T-connector to the scope, triggering on the pulser signal. We saw now signals --> lower the threshold of the fast comparator to 1 MeV - 0x64 on all ASICs in aida10. 677/10

No signal was observed, threshold was then set to 0x190 (400 keV) 677/11

13:00 We resumed after lunch. It seems that our thresholds were way too high for starters each channel was actually 100 keV in HEX. We then set the thresholds in aida10 to 0x32, 0x20, 0x20, 0x11 for ASIC 1,2,3,4, respectively. This was to achieve a 10 Hz hit rate in all of the channels! This was done successfully see 677/12 for the hit rate spectrum and also 677/13-16 for the thresholds.

 

 

The pulser was set to 0.5 V to test if we can still see the time spectrum between AIDA and bPlast with reduced thresholds -- mimicking a beta event.

These thresholds were changes from 0x32, 0x20, 0x20, 0x11 for ASIC 1,2,3,4 ---> 0x10, 0x10, 0x0d, 0x10.

This was set to have the HitRate in aida10 to be just above the noise.

Attachment 1: Screenshot_from_2025-01-08_17-51-51.png  130 kB  | Hide | Hide all
Screenshot_from_2025-01-08_17-51-51.png
ELOG V3.1.4-unknown