AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
DESPEC |
 |
|
Message ID: 690
Entry time: Sat Jan 11 14:05:03 2025
|
Author: |
TD |
Subject: |
Saturday 11 January 2024 |
|
|
14.37 Detector bias & leakage current OK - attachment 1
FEE64 temperatures OK - attachment 2
WR timestamps - attachment 3
aida06 out of sync
global clock status - attachment 4
aida06 fails
ADC calibration - attachment 5
all fail - probably due to ASIC synchronise *after* ADC calibration
WR decoder & FPGA TS status - attachments 6-7
aida06 status 0xd
DAQ run control - attachment 8
aida06 undefined => aida06 has rebooted
Merger & Tape Server - attachments 9-10
merger, tape server and data transfer have stopped
See https://elog.ph.ed.ac.uk/DESPEC/691
aidao06 system console - aida06 rebooted sometime yesterday
*** Recovery
For FEE64 aida06 *only*
DAQ reset
DAQ setup
For all FEE64s
DAQ stop
Restart merger, merger setup, merger go
Following two merger restart cycles all FEE64s stopped OK
Recovery complete
All system wide checks OK
Disable data transfer, DAQ go
per FEE64 rate spectra = attachment 11
aida04 high rates
per FEE64 1.8.W spectra - 20us FSR - attachments 12-13
DAQ stop, enable data transfer, DAQ go
Merger and server consoles - attachment 14
merger and tape server (no storage mode) running OK but no data transfer to MBS server
WR timestamps OK - attachment 15
ADC and DISC data item stats - attachment 16-17
aida04 & aida16 high rates - all others appear to be OK
ASIC check load for all aida16 ASICs changes rate from c. 420k to 0
ASIC check load for all aida04 ASICs changes rate from c. 200k to c. 120k
18.00 No data transfer to MBS - nothing useful being donme at the moment - will place system in safe state
DAQ stop
Detector bias OFF
FEE64 power OFF |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|