Message ID: 4     Entry time: Wed May 15 06:40:16 2019
Author: Algora, Kiss, Nacher  
Subject: DTAS aligned and calibrated 

Signal cables (long cables, from preamp to digitizer) of channels 13 and 16 were changed. For the moment they have not been relabelled.

We assumed that it was not a HV supply or cable problem, since the anods were aligned and only the dynodes moved.

After changing the cables we realigned the dynodes and recalibrated all signals. Screen shots are attached.

File (root file) with the two sources (Co and Cs) saved as 201905151432_137Cs60Co.root

Configuration file: 20190515_Conf.xlsx

Rate with two sources: 52 kHz (Dynodes), 47 kHz (Anodes)

 