Signal cables (long cables, from preamp to digitizer) of channels 13 and 16 were changed. For the moment they have not been relabelled.
We assumed that it was not a HV supply or cable problem, since the anods were aligned and only the dynodes moved.
After changing the cables we realigned the dynodes and recalibrated all signals. Screen shots are attached.
File (root file) with the two sources (Co and Cs) saved as 201905151432_137Cs60Co.root
Configuration file: 20190515_Conf.xlsx
Rate with two sources: 52 kHz (Dynodes), 47 kHz (Anodes)
|