AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DTAS  ELOG logo
Message ID: 67     Entry time: Wed May 29 07:04:19 2019
Author: A. Algora 
Subject: F11 plastic connected 

F11 plastic has been connected to our dacq as it was in the earlier BRIKEN setting.

The detector cables were followed and connections checked.

The outputs of the detector (L and R) go through an Ortec TFA 474 with the following settings:

Coarse gain: x4

Fine gain: looks minimal

Integrate (ns): 200

Diff (ns): 100

Non-inv setting

After the TFA, the Right signal is plugged in channel 1 of our card, and an attenuated signal (50 Ohm) is

plugged in channel 3. Similarly the Left signal after the TFA is plugged in channel 2 and the attenuated 

signal is plugged in channel 4.


Parameters of the channels of the dacq have to be checked.

ELOG V3.1.4-unknown