  10   Tue Jul 9 13:31:20 2019 CBTests with positive pulsesAccording to Carl U. the DPP-PSA software
is not meant to be used with negative pulses,
and some of the
 190709_PositivePulse.png190709_BaselineMean.png 
  9   Tue Jul 9 13:05:17 2019 CBJul 8Found VME crate yesterday reporting AC power
failure. Workstation unresponsive.
  
  8   Fri Jun 28 18:01:09 2019 CBLong run to test DAQ stabilityLeft DAQ running with pulser signal to attempt
to reproduce error leading loss of communication
with VME crate.
  
  7   Wed Jun 26 17:04:07 2019 CBNew settings for internal CFDTried using settings attached to enable internal
CFD.
Restored / DAQ Go. No waveforms.
 1725_PosSig_Working.txt 
  6   Wed Jun 26 14:19:36 2019 CBAnalysis of R2Analysed first ~500 MB of R2_0 using sorting
program
 260619_R2Histo.png 
  5   Wed Jun 26 14:15:43 2019 CBWaves to histogram programWrote simple program to convert from waves
to amplitude/energy histogram.
Program uses semi-gaussian filtering with
 WavesToHisto.cpp 
  4   Tue Jun 25 11:51:00 2019 CBTest run 4P9 3-alpha source
Hamamatsu Photodiode - bias = 15V
Cooknell preamplifier
  
  3   Mon Jun 10 06:17:51 2019 CB, RG, AK3-6 June - first DAQ testsTested performance of DAQ with following equipment

Hamamatsu photodiode (S3590-09)
  
  2   Mon Jun 10 06:09:59 2019 CB, OHSorting programSort program written by Oscar, adapted from
the RIKEN AIDA sort.
EDIT 26 Jun 2019 by CB - C++11 dependencies
 dataReader.tar.gz 
  1   Mon Jun 10 03:26:41 2019 CBDAQ setup, start & stop instructions(Updated to 10/01/2020)

If recovering from a crash, turn off the
 7x 
ELOG V3.1.4-unknown