Message ID: 99     Entry time: Wed Aug 6 11:20:52 2025
Author: Selin 
Subject: Pulser Test 

The pulser test was performed with a frequency of 500 Hz and with voltage settings of: 2V, 1.6V, 1.2V, 0.8V, 0.6V, 0.4V, 0.3V, 0.2V, 0.1V, 0.05V, 0.025V.

The run was started with calibration settings and then the voltage was changed every 2.5-3 minutes.

From the attached plots, it can be seen that everything is as expected. This means that the pulser multipolarity is consistent with front and back strips, pulser signals are more frequent and the pulser amplitude decreases with decreasing V.

The front and back pulser tests have been done with separate runs and the settings have been reverted to their original values (100 Hz, 2V, pulser is off).

Attachment 1: front_2V.png  39 kB  | Hide | Hide all
front_2V.png
Attachment 2: front_less2V.png  41 kB  | Hide | Hide all
front_less2V.png
Attachment 3: front_less1V.png  54 kB  | Hide | Hide all
front_less1V.png
Attachment 4: back_2V.png  44 kB  | Hide | Hide all
back_2V.png