Message ID: 236
Entry time: Wed May 25 18:51:54 2016
Reply to this: 237
|
Author: |
Patrick Coleman-Smith |
Subject: |
[HowTo] Use the Multiplicity Trigger Firmware, V8.18, and update the files |
|
|
The attached firmware file, FEE_GF_Feb16_18.bin, should be saved to /MIDAS/Aida and the FlashPgm.csh file edited
to load "FEE_GF_Feb16_18.bin"
Use the Run Control Expert command to update the firmware on all the FEE64s.
Power cycle the FEE64s to load the new firmware.
Make a backup copy of the files in the directory /MIDAS/TclHttpd/Html/AIDA/DISC.
Unzip the files from DISC_25May16.zip and copy them into /MIDAS/TclHttpd/Html/AIDA/DISC to replace the existing
ones.
Make a backup copy of the file /MIDAS/TclHttpd/Html/AIDA/LOCAL.tml
Unzip the file from LOCAL_tml_25May16.zip and copy into /MIDAS/TclHttpd/Html/AIDA/LOCAL to replace the existing
file.
'RESET' and browser refresh the two updated control windows to load the new software.
-----------------------------------------------------
To use the Multiplicity Trigger:-
In the Local Controls select option 5 for the Trigger.
In the Discriminator window.
Set the 'Upper Limit'
Set the 'Lower Limit' ( always greater than 0 ..... or the Trigger will always be present )
Set the 'Time window'. The number of 10nS clocks. So 4 => 40nS. Register is 0 to 255.
-----------------------------------------------------
These registers are not in the Save/Restore set yet.
Tested on the T9 system using the pulser and setting/clearing Mask bits.
Please let me know how this works for you. |
|
|
|