|
AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
|
Message ID: 237
Entry time: Thu May 26 01:44:45 2016
In reply to: 236
|
Author: |
Patrick Coleman-Smith |
Subject: |
[HowTo] Use the Multiplicity Trigger Firmware, V8.18, and update the files |
|
|
Done. Awaiting test. See attachments 1-2.
> The attached firmware file, FEE_GF_Feb16_18.bin, should be saved to /MIDAS/Aida and the FlashPgm.csh file edited
> to load "FEE_GF_Feb16_18.bin"
>
> Use the Run Control Expert command to update the firmware on all the FEE64s.
>
> Power cycle the FEE64s to load the new firmware.
>
> Make a backup copy of the files in the directory /MIDAS/TclHttpd/Html/AIDA/DISC.
>
> Unzip the files from DISC_25May16.zip and copy them into /MIDAS/TclHttpd/Html/AIDA/DISC to replace the existing
> ones.
>
> Make a backup copy of the file /MIDAS/TclHttpd/Html/AIDA/LOCAL.tml
>
> Unzip the file from LOCAL_tml_25May16.zip and copy into /MIDAS/TclHttpd/Html/AIDA/LOCAL to replace the existing
> file.
>
> 'RESET' and browser refresh the two updated control windows to load the new software.
> -----------------------------------------------------
>
> To use the Multiplicity Trigger:-
> In the Local Controls select option 5 for the Trigger.
>
> In the Discriminator window.
>
> Set the 'Upper Limit'
> Set the 'Lower Limit' ( always greater than 0 ..... or the Trigger will always be present )
> Set the 'Time window'. The number of 10nS clocks. So 4 => 40nS. Register is 0 to 255.
>
> -----------------------------------------------------
>
> These registers are not in the Save/Restore set yet.
>
> Tested on the T9 system using the pulser and setting/clearing Mask bits.
>
> Please let me know how this works for you. |
|
|
|
|