Message ID: 381
Entry time: Mon Oct 10 03:15:59 2016
|
Author: |
CG & TD |
Subject: |
Monday 10 October |
|
|
10.30 DAQ found stopped
All FEE64 AIDA SYNC rates zero, AIDA SYNC count ~ 4.4M => DAQ stopped @ 01.40
3He detector DAQ off - no 50MHz clock to Master MACB
System wide checks - see attachments 26-18
N.B. It would appear that the DAQ stalled (TS_SYNC_PHASE 1) c. 01.40 and at some
later time (c. 06.00) the 3He detector DAQ was powered off which means that AIDA
lost the external 50MHz clock to the Master MACB - system-wide screenshots should
be interpreted with caution.
14.15 3He detector DAQ 50MHz clock restarted
VME and NIM crates ON
Excecute command ConfClock_50MHz on 3He detector DAQ computer
Power cycle AIDA FEE64s
Set TS_SYNC_PHASE 2
DSSSD bias/leakage current OK
FEE64 temperatures OK
*All* system-wide checks OK
B2F/F11 temperature + deg C, d.p. + deg C,RH %
ASIC settings 2016Sep02-18.17.01
slow comparator 0x10 -> 0x0a
Master MACB serial #13 (see https://elog.ph.ed.ac.uk/AIDA/367)
Master MACB mode 5
14.20 DAQ starts (no data transfer)
18.10 DAQ found stopped
All FEE64 AIDA SYNC rates zero, AIDA SYNC count ~ 3.4M => DAQ stopped @ 16.50
System-wide checks - see attachments 16-9
Power cycle AIDA FEE64s
Set TS_SYNC_PHASE 3
ASIC settings 2016Sep02-18.17.01
slow comparator 0x10 -> 0x0a
20.10 DAQ found stalled
System-wide checks - see attachments 8-1
Power cycle AIDA FEE64s
Set TS_SYNC_PHASE 0
ASIC settings 2016Sep02-18.17.01
slow comparator 0x10 -> 0x0a
20.35 DAQ starts (no data transfer)
slow comparator 0x10 -> 0x0a
18.35 DAQ starts (no data transfer)
20.16 DAQ found stopped
All FEE64 AIDA SYNC rates zero, AIDA SYNC count ~ 120k (17/24 FEE64s) or 1.49M (7/24) => FEE64s stopped @ 18.40 & 19.40
20.36 DAQ starts (no data transfer)
23.47 DAQ OK |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|