AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  AIDA  ELOG logo
Message ID: 408     Entry time: Sun Oct 30 01:53:13 2016
Author: TD 
Subject: Sunday 30 October 
10.53 DAQ found stalled 
      Most recent files R89_102 last accessed 07.17

      DSSSD bias & leakage current OK - see attachments 1 & 2
      FEE64 temperatures OK - see attachment 3

      All system wide checks OK *except* Clock status, SYNC errors & SYNC pulses - see attachments 4-8 - unusual
messages/reports

      One merger error messages - link 18 - see attachment 9

      Restart merger
      DAQ STOP OK
      DAQ RESET & SETUP OK
      ReSYNC OK
      All system-wide checks OK
      Disable H & data transfer #1
      Restart merger & tapeserver
      Enable H & data transfer #1

      Standard soft reset did not work - one or more FEE64s dropped data transfer #1 link at DAQ GO

      Retry - one, more FEE64s indicate error state - run control tab unresponsive 

      Power cycle FEE64s 
Attachment 1: 1.png  12 kB  | Hide | Hide all
1.png
Attachment 2: 2.png  12 kB  | Hide | Hide all
2.png
Attachment 3: 3.png  135 kB  | Hide | Hide all
3.png
Attachment 4: 4.png  206 kB  | Hide | Hide all
4.png
Attachment 5: 5.png  162 kB  | Hide | Hide all
5.png
Attachment 6: 6.png  162 kB  | Hide | Hide all
6.png
Attachment 7: 7.png  218 kB  | Hide | Hide all
7.png
Attachment 8: 8.png  218 kB  | Hide | Hide all
8.png
Attachment 9: 9.png  89 kB  | Hide | Hide all
9.png