Message ID: 408
Entry time: Sun Oct 30 01:53:13 2016
|
Author: |
TD |
Subject: |
Sunday 30 October |
|
|
10.53 DAQ found stalled
Most recent files R89_102 last accessed 07.17
DSSSD bias & leakage current OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
All system wide checks OK *except* Clock status, SYNC errors & SYNC pulses - see attachments 4-8 - unusual
messages/reports
One merger error messages - link 18 - see attachment 9
Restart merger
DAQ STOP OK
DAQ RESET & SETUP OK
ReSYNC OK
All system-wide checks OK
Disable H & data transfer #1
Restart merger & tapeserver
Enable H & data transfer #1
Standard soft reset did not work - one or more FEE64s dropped data transfer #1 link at DAQ GO
Retry - one, more FEE64s indicate error state - run control tab unresponsive