09.13 DAQ run R3 ends OK (final file R3_9)
ASIC settings 2017Mar27-16.12.39
slow comparator 100 (dec)
LEC/MEC fast comparator 255 (dec)
BNC PB-4 pulser OFF
09.20 Biases & leakage currents OK - see attachments 1 & 2
Ambient temperature +23.8 deg C, d.p. -3.5 deg C, RH 15.7%
Julabo FL11006 set point +20 deg C actual +20.0 deg C
09.32 FEE64 temperatures OK - see attachment 3
09.39 System-wide checks OK except
ADC calibration fails nnaida21 & 23
SYNC errors not counted since baseline
SYNC pulses received
see attachments 4-8
13.36 PE16/R1
Pulser walkthough
BNC PB-4 settings
Amplitude 90,000 - 10,000 @ 10,000 step
x5 attenuator IN
t_r 50ns, tau_d 50us
polarity -
rate ~250Hz
delay MIN, tail pulse
ASIC settings 2017Mar27-16.12.39
see attachment 9
13.52 PE16/R2
Pulser walkthough
BNC PB-4 settings
Amplitude 10,000 - 2,000 @ 2,000 step
see attachment 10
14.03 PE16/R3
Pulser walkthough
BNC PB-4 settings
Amplitude 2,000 - 8,000 @ 2,000 step, 10,000 - 50,000 @ 10,000 step
see attachment 11
20.30 System-wide checks
All OK *except* clock status - nnaida1 global clocks failed 6 - see attachment 12
20.35 PE16/R4
Background
BNC PB-4 OFF
ASIC settings 2017Mar27-16.12.39
MACB master mode 5 (external 50MHz clock from Tektronix AFG3102C)
20.40 Biases & leakage currents OK - see attachments 13 & 14
Ambient temperature +23.6 deg C, d.p. -3.2 deg C, RH 16.6%
Julabo FL11006 set point +20 deg C actual +20.0 deg C
FEE64 temperatures OK - see attachment 15
20.50 PE16/R4 ends OK |