> 10.35 DAQ continues OK
> Current file May2017/R19_253
>
> ASIC settings 2017Mar27-16.12.39 *except*
> slow comparator 0x0a
> LEC/MEC fast comparator 0xff
>
> Detector biases; leakage currents OK - see attachment 1
> FEE64 Temperatures OK - see attachment 2
> good events; statistics OK - see attachment 3
> No warnings/errors reported by merger since DAQ GO - see attachment 4
>
> *all* system wide checks OK *except* ADC calibration of nnaida7
>
> 10.38 E17/F11 ambient temperature +24.7 deg C, RH 40.7%, d.p. +10.7 deg C
> Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 75%)
>
> 10.39 DAQ stops OK
> Current file May2017/R19_253
> Deleted R19*
>
> ASIC settings 2017Mar27-16.12.39 *except*
> slow comparator 0x0a
> LEC/MEC fast comparator 0xff
>
> BNC PB-4 settings
> Amplitude 90,000
> x5 attenuator IN
> t_r 50ns, tau_d 50us
> polarity -
> rate 250Hz range
> delay MIN, tail pulse
>
> Check ASIC Control x2
> Zero all spectra
>
> 10.47 DAQ starts
> File May2017/R20
> Ras Pi system console logs for USB0 attached - see attachment 5
> Analysed R20_0. All appears to be working fine. Highest dead time ~0.3% - see attachment 6
> Off-line analysis of R20_0 confirms ADC data and discriminator are synchronised - see attachments 7,8,9
Concerning nnaida7 and the failure of calibration..... have you tried to calibrate it using the FADC Align & Control ?
If it still fails then there is a question about the quality of the clock signal being received by this module.
I will log through and look at the console log for any anomolies but please try and calibrate.
Let me know if you wish me to take remote control. |