AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  AIDA  ELOG logo
Entry  Mon May 22 02:34:51 2017, CB, TD, Monday 22 May 16x
    Reply  Mon May 22 08:43:10 2017, CB, TD, Patrick, Monday 22 May 
       Reply  Mon May 22 08:53:42 2017, CB, TD, Patrick, Monday 22 May 
Message ID: 585     Entry time: Mon May 22 08:53:42 2017     In reply to: 584
Author: CB, TD, Patrick 
Subject: Monday 22 May 
> Concerning nnaida7 and the failure of calibration..... have you tried to calibrate it using the FADC Align  & Control ?
> If it still fails then there is a question about the quality of the clock signal being received by this module.
> I will log through and look at the console log for any anomolies but please try and calibrate.
> Let me know if you wish me to take remote control.

Yes - there have been multiple attempts to re-calibrate nnaida7 ADCs

Note *all* ADCs were calibrated before the most recent FEE64 power cycle.

It is not unusual for ~1-3 FEE64s to fail ADC calibration tests following
a power cycle - there is no obvious pattern to which ADCs fail or which (if
any) will be successfully re-calibrated.