13.38 DAQ continues OK
Current file May2017/R27_49
Detector biases; leakage currents OK - see attachment 1
FEE64 Temperatures OK - see attachment 2
good events; statistics OK - see attachment 3
No warnings/errors reported by merger since DAQ GO - see attachment 4
*all* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (=6)
10.30 E17/F11 ambient temperature +23.8 deg C, RH 40.9%, d.p. +9.6 deg C
Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 75%)
13:40 Stopped DAQ running on alpha background calibration run
Current file May2017/R49
tyUSB7 log (nnaida1) attached
13:43 Detector bias OFF
14.16 DAQ starts
File May2017/R28
Pulser walkthrough
BNC PB-4
Amplitude 90,000-10,000 @ 10,000 step
x5 attenuator IN
t_r 100ns tau_d 50us
- polarity
frequency 250Hz range
delay MIN, tail pulse, INT ref
ASIC settings 2017Mar27-16.12.39 *except*
slow comparator 0x0a
LEC/MEC fast comparator 0x20
See attachment 6
14.25 DAQ stop OK
14.27 DAQ starts
File May2017/R29
Pulser walkthrough
BNC PB-4
Amplitude 12,000-2,000 @ 2,000 step
ASIC settings 2017Mar27-16.12.39 *except*
slow comparator 0x0a
LEC/MEC fast comparator 0x20
See attachment 7
14.34 DAQ stop OK
14.38 DAQ starts
File May2017/R30
Background, pulser OFF
ASIC settings 2017Mar27-16.12.39 *except*
slow comparator 0x0a
LEC/MEC fast comparator 0x20
15.16 DAQ stop OK
May2017/R30_8
16:17 DAQ starts okay
current run May2017/R31_0
leakage current and temperature fine
see attachment 8 and 9
|