AIDA |
 |
|
Message ID: 634
Entry time: Mon Jun 5 16:02:50 2017
|
Author: |
OH, PV, AE |
Subject: |
Tuesday 06 June 00.00-08.00 |
|
|
00:03 BigRIPS starts run 3007
BRIKEN starts run 007
AIDA on file R5_146
00:07 Bias and leakage currents ok - attachment 1
System wide checks all ok
FEE temperatures ok - attachment 2
Good evts still seeing noise on 13 and 14 - attachment 3
No merger errors since GO
00:48 DSSD Rates from R5_160
*** DSSSD # 1 count: 250904 old count: 163089 dt: 1.02 s LEC rate: 86326.91 Hz
*** DSSSD # 2 count: 335185 old count: 222784 dt: 1.02 s LEC rate: 110496.28 Hz
*** DSSSD # 3 count: 157445 old count: 103991 dt: 1.02 s LEC rate: 52548.18 Hz
*** DSSSD # 4 count: 102079 old count: 67878 dt: 1.02 s LEC rate: 33621.44 Hz
*** DSSSD # 5 count: 40500 old count: 26651 dt: 1.02 s LEC rate: 13614.32 Hz
*** DSSSD # 6 count: 64652 old count: 42396 dt: 1.02 s LEC rate: 21878.85 Hz
*** DSSSD # 1 count: 0 old count: 0 dt: 1.02 s HEC rate: 0.00 Hz
*** DSSSD # 2 count: 41 old count: 27 dt: 1.02 s HEC rate: 13.76 Hz
*** DSSSD # 3 count: 161 old count: 103 dt: 1.02 s HEC rate: 57.02 Hz
*** DSSSD # 4 count: 131 old count: 85 dt: 1.02 s HEC rate: 45.22 Hz
*** DSSSD # 5 count: 22 old count: 12 dt: 1.02 s HEC rate: 9.83 Hz
*** DSSSD # 6 count: 9 old count: 6 dt: 1.02 s HEC rate: 2.95 Hz
01:02 Briken and BigRIPS daq stops
01:03 Briken Starts run 8
BigRIPS starts run 3008
01:40 Phong noticed a significant difference in the noise in the (raw) low energy spectra a few of the FEEs when
comparing the runs right before and right after the power-cycle (See Attachment 12) required to recover
the DAQ crash last afternoon (June 5th).
NNAIDA 22, 21, 14, and 13 had a much higher count rate after the power cycle, and what seems like a lower
threshold. We can't observe the pulser peak in any of those.
NNAIDA 13 has the highest count rate, and the waveforms show what looks like noise with a regular
frequency (attachments 4 - 7)
For reference, we include the waveform of other positive FEEs 8-11
02:51 Bias and leakages ok - attachment 13
System wide checks ok
Good Event statistics - attachment 14
Fee Temperatures - attachment 15
03:03 BigRIPS and BRIKEN stop daqs
AIDA on file R5_208
03:04 BigRIPS starts run 3009
BRIKEN starts run 9
AIDA on file R5_208
04:03 BigRIPS and BRIKEN stop daqs
AIDA on file R5_230
04:04 BigRIPS starts run 3010
BRIKEN starts run 10
AIDA on file R5_230
04:09 System wide checks all ok
Bias and leakages ok - attachment 16
Fee Temperatures - attachment 17
Good Event Statistics - attachment 18
04:48 DSSD rates from R5_244
*** DSSSD # 1 count: 1239207 old count: 1155424 dt: 1.04 s LEC rate: 80379.42 Hz
*** DSSSD # 2 count: 1684042 old count: 1567549 dt: 1.04 s LEC rate: 111760.61 Hz
*** DSSSD # 3 count: 806937 old count: 752064 dt: 1.04 s LEC rate: 52643.85 Hz
*** DSSSD # 4 count: 505814 old count: 472642 dt: 1.04 s LEC rate: 31824.43 Hz
*** DSSSD # 5 count: 208086 old count: 193231 dt: 1.04 s LEC rate: 14251.53 Hz
*** DSSSD # 6 count: 335686 old count: 313190 dt: 1.04 s LEC rate: 21582.13 Hz
*** DSSSD # 1 count: 8 old count: 7 dt: 1.04 s HEC rate: 0.96 Hz
*** DSSSD # 2 count: 193 old count: 181 dt: 1.04 s HEC rate: 11.51 Hz
*** DSSSD # 3 count: 872 old count: 810 dt: 1.04 s HEC rate: 59.48 Hz
*** DSSSD # 4 count: 637 old count: 601 dt: 1.04 s HEC rate: 34.54 Hz
*** DSSSD # 5 count: 89 old count: 81 dt: 1.04 s HEC rate: 7.68 Hz
*** DSSSD # 6 count: 71 old count: 65 dt: 1.04 s HEC rate: 5.76 Hz
05:01 BRIKEN and BigRIPS stop DAQs
AIDA on file R5_250
05:02 BigRips starts run 3011
BRIKEN starts run 11
AIDA on file R5_250
06:02 BigRIPS and BRIKEN stop their daqs
AIDA on file R5_272
06:03 BigRIPS starts run 3012
BRIKEN starts run 12
AIDA on file R5_272
06:06 System wide checks all ok
Bias and leakage currents ok - attachment 19
Good event statistics - attachment 20
FEE temperatures ok - attachment 21
06:16 Analysis of R5_270 - attachment 22
Three MBS errors
06:34 130Cd decay curve - attachments 22 and 23
Despite the large amounts of noise in the detectors it is still possible to observe decay curves. They are
cleaned up a lot
by applying a multiplicity cut (attachment 22 with attachment 23 without. Further cuts and vetos could be
applied to clean up further.
The fit in attachment 22 gives a half life of 182ms which is not quite the expected 162ms but a reasonable
first fit.
07.14 Layout - 1.8.W waveforms - see attachments 24-27
No significant change in intrinsic/extrinsic noise cf. previous screenshots
nnaida13 ASIC 1-4 waveforms - see attachments 28-31 - range 100*20us
07.24 BRIKEN and BigRIPS stop DAQs
AIDA on file R5_300
07.25 Stop AIDA DAQ
Cycle detector biases
Start AIDA DAQ
07.38 BigRips starts run 3014
BRIKEN starts run 14
AIDA on file R6_1
07.45 nnaida13 ASIC 1-4 waveforms - see attachments 32-35 - range 100*20us
08.07 all system wide checks OK |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|