AIDA |
 |
|
Message ID: 687
Entry time: Sat Oct 28 03:33:10 2017
|
Author: |
TD, OH |
Subject: |
Saturday 28 October |
|
|
11.33 DAQ running OK - date transfer not enabled
BNC PB-4
Amplitude 90,000
x5 attenuator IN
t_r 100ns tau_d 50us
- polarity
frequency range 25Hz
delay MIN, tail pulse, INT ref
ASIC settings 2017Oct23-14.38.03
*all* FEE64s slow comparators 0xa *except* nnaida13 0xe, nnaida19-22 0xc
Detector biases & leakage currents OK - see attachments 1 & 2
FEE64 Temperatures OK - see attachment 3
DAQ statistics OK - see attachment 4
*all* system wide checks OK *except* nnaida21 fails ADC calibration
Julabo FL11006 recirculating chiller set point +20 deg C, actual +20.0 deg C
Ambient temperature +26.2 deg C, d.p. +6.2 deg C, RH 28.3%
14.40 Detector biases & leakage currents OK - see attachments 5 & 6
FEE64 Temperatures OK - see attachment 7
DAQ statistics OK - see attachment 8
*all* system wide checks OK *except* nnaida21 fails ADC calibration
14.50 DAQ starts
AIDA data file Oct17/R18
14.53 per FEE64 rates (20us/channel) - see attachment 9
No indication of c. 400us period noise as indicated by DAQ statistics attachment 8
15.46 DAQ stopped
DAQ start - data tranfser disabled
19.40 Detector biases & leakage currents OK - see attachments 10 & 11
FEE64 Temperatures OK - see attachment 12
DAQ statistics OK - see attachment 13
*all* system wide checks OK *except* nnaida21 fails ADC calibration |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|