AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 690
Entry time: Sun Oct 29 05:16:35 2017
|
Author: |
TD, OH |
Subject: |
Sunday 29 October |
|
|
14.16 DAQ running OK - data transfer not enabled
BNC PB-4
Amplitude 90,000
x5 attenuator IN
t_r 100ns tau_d 50us
- polarity
frequency range 25Hz
delay MIN, tail pulse, INT ref
ASIC settings 2017Oct23-14.38.03
*all* FEE64s slow comparators 0xa *except* nnaida13 0xe, nnaida19-22 0xc
Detector biases & leakage currents OK - see attachments 1 & 2
FEE64 Temperatures OK - see attachment 3
DAQ statistics OK - see attachment 4
*all* system wide checks OK *except* nnaida21 fails ADC calibration
Julabo FL11006 recirculating chiller set point +20 deg C, actual +20.0 deg C
Ambient temperature +23.9 deg C, d.p. +10.8 deg C, RH 43.2%
15.05 per FEE64 Rate, 1.8.L and 1.8.W spectra - note common x & y scales
Peak width nnaida5 63 ch FWHM
Peak width nnaida8 82 ch FWHM
15.09 All histograms zero'd
18.25 Detector biases & leakage currents OK - see attachments 11 & 12
FEE64 Temperatures OK - see attachment 13
DAQ statistics OK - see attachment 14
*all* system wide checks OK *except* nnaida21 fails ADC calibration |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|