Message ID: 760
Entry time: Sat Nov 3 00:37:29 2018
|
Author: |
CA, CB, TD, OH |
Subject: |
Saturday 3 November |
|
|
09.37 DAQ continues OK - data transfer disabled
FEE64 LEC/MEC slow comparator threshold 0xa
09.39 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK *except* nnaida17 c. 400k - see attachment 4
nnaida2 rate due to one hot channel as before
1.8.L spectra - see attachments 5 & 6
N.B. evidence of minimally ionising particles (presumably protons) from beam to F11 E detector
~10-15cm Pb bricks were placed within BRIKEN (CH2)n stack to stop heavy particles
1.8.W spectra 20us FSR - see attachments 7 & 8
10.10 LEC/MEC slow comparator from 0xa to 0x64
3x ASIC check load
DAQ stop
all histograms zero'd
enable data transfer
DAQ start - file Oct18/R16
beam background
good event statistics OK *except* nnaida2 20kHz - see attachment 9 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|