AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 767
Entry time: Tue Nov 6 00:13:00 2018
|
Author: |
CB. PJW, TD, OH, CA |
Subject: |
Tuesday 6 November |
|
|
09:13 DAQ still running R18_1
Merger and TapeServer OK
System wide checks ok, except for nnaida7 ADC calibration failure
Temperature OK (attachment 1)
Bias and leakage current OK (attachment 2)
Good event statistics OK (attachment 3)
11.20 1.8.W spectra - attachment 4 & 5
analysis file Oct18/R18_1 - attachment 6
15:35 Inserted AIDA snout in BRIKEN. Lowered threshold to 100 keV. Rate looks OK.
R19 pulser walkthrough 90 0000 ... 10 0000 (10 000 steps) - attach 7&8
R21 pulser walkthrough 12 000 ... 2 000 (2 000 steps)
Pulser back to 90 000. Frequency set to zero.
R22 background run starts.
Rates OK (attach 9)
16.07 Raised slow comparator threshold to 100 from 10
DAQ start (file R23)
Merger and TapeServer ok, no error messages in terminal
Good event statistics ok (attachment 10)
System wide checks ok, except nnaida7 failing ADC calibration
FEE64 temperatures ok (attachment 11)
bias/leakage currents ok (attachment 12)
16.58 Repeat pulser walkthroughs with additional ASIC check loads
R25 pulser walkthrough 90 0000 ... 10 0000 (10 000 steps) - attach 7&8
R26 pulser walkthrough 12 000 ... 2 000 (2 000 steps)
18.26 DAQ start - file Oct18/R27
alpha background
LEC/MEC slow comparator 0xa -> 0x64
BNC PB-4 pulser OFF @ 18.29!
18.34 1.8.W spectra 20us FSR - see attachments 13 & 14 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|