AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 770
Entry time: Thu Nov 8 23:35:50 2018
|
Author: |
OH, TD, CA, CB |
Subject: |
Friday 9 November |
|
|
08:36 DAQ still running
Not writing data to disk
Bias and leakage currents ok - attachment 1
System wide checks all ok
FEE Temperatures ok - attachment 2
Statistics - attachment 3
Check ASIC Control done
08:56 Pulser peaks are ok and similar to what has been observed over the past two weeks
Layout 4 and 5 - attachments 4 and 5
11.55 Lost connection to FEE64 1-16 - fuse blown
stopped DAQ
switched all FEE64 off
Replaced 20A fuse
switched FEE64 back on individually
system wide checks ok, except nnaida7 failure in ADC calibration
DAQ start (not writing to file)
FEE64 temperatures ok (attachment 6)
Bias/Leakage currents ok (attachment 7)
good event statistics ok (attachment 8)
17.08 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 9 & 10
FEE64 temperatures OK - see attachment 11
good event statistics OK - see attachment 12
per FEE64 rate spectra - see attcahment 13
1.8.L spectra - see attachments 14 & 15
nnaida1 pulser peak width 72 ch FWHM
1.8.W spectra 20us FSR - see attachments 16 & 20
1.8.W spectra 200us FSR - see attachments 17 & 21
1.8.W spectra 2ms FSR - see attachments 18 & 22
1.8.W spectra 20ms FSR - see attachments 19 & 23 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|