AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 795
Entry time: Mon May 13 03:23:58 2019
|
Author: |
CA, TD, OH |
Subject: |
Monday 13 May |
|
|
09.55 all system wide checks OK *except* nnaida7 fails ADC calibration
FEE64 temperatures OK - see attachment 1
good event statistics OK - see attachment 2
all FEE64s <100k, 11x < 20k
1.8.L spectra - see attachments 3 & 4
nnaida1 70ch FWHM, nnaida3 80 ch FWHM
detector biases & leakage currents OK - see attachments 5 & 6
1.8.W spectra 20us FSR - see attachments 7 & 8
nnaida1 Options, Local Controls, Data Redaout, Discriminator Controls - see attachment 9-13
Lost Activity Monitor -see attachment 14
18.01 DAQ found stopped
This was caused by a chnage to the AIDA external clock frequency - see https://elog.ph.ed.ac.uk/AIDA/798
Following change of AIDA external clock frequency to 50MHz
DAQ RESET
DAQ SETUP
Sync all ASIC clocks using Resync
all system wide checks OK *except* nnaida7 fails ADC calibration
zero all histograms
DAQ GO
18.31 FEE64 temperatures OK - see attachment 15
good event statistics OK - see attachment 16
all FEE64s <100k, 10x < 20k
detector biases & leakage currents OK - see attachments 17 & 18 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|