AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 799
Entry time: Tue May 14 02:19:20 2019
|
Author: |
CA, OH, TD |
Subject: |
Tuesday 14th May 2019 |
|
|
10.20 System wide checks ok (except naida07 ADC calibration failure)
FEE temperatures ok (attachment 1)
good event statistics ok (attachment 2)
detector bias and leakage currents ok (attachment 3)
DAQ stopped
10.40 pulser switched off
slow comparator threshold raised to 0x64
DAQ, merger, TapeServer started - all ok (attachments 4 and 5)
writing to file R1 in TapeData/May19 at 70 kb/s
good event statistics (attachment 6)
16:11 Tapeserver and merger still ok
DAQ still running
System wide checks all ok *except nnaida7 fails ADC calibration*
Bias and leakage currents ok - attachment 7
FEE Temperatures ok - attachment 8
Statistics still ok
Spectrum are showing alpha events (Layout 4) - attachment 9
18:25 Tektronix clock accidentally unplugged. DAQ has stalled as a result.
Clock reset to 50MHz 5V TTL signal
18:29 System wide check of clock status shows error message in attachment 10
DAQ Reset
With slow comp at oxa statistics are still good - attachment 11
System wide checks all ok *except nnaida7 fails ADC calibration*
Resync to the ASIC clock
All histograms zeroed
18:37 Slow comp threshold raised to 0x64 for alpha run
Writing to file:
May19/R2_0
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|