AIDA |
 |
|
Message ID: 802
Entry time: Fri May 17 02:15:56 2019
|
Author: |
CA, TD, OH |
Subject: |
Friday 17 May |
|
|
10.18 DAQ continues OK to file May19/R5_1
alpha background
ASIC settings 2018Oct03-13.26.58
*except* slow comparator 0xa -> 0x64
BNC PB-4 Pulser OFF
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachments 4
Merger OK - see attachment 5
TapeServer OK - see attachment 6
1.8.W spectra 20us FSR - attachments 7 & 8
1.8.W spectra 20ms FSR - attachments 9 & 10
Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C - water topped up
Temperature +24.8 deg C dew point +8.3 deg C RH 35.3%
10.42 DAQ stop May19/R5_2
slow comparator 0x64 -> 0xa
Check ASIC Control *all* ASICs & FEE64s
10:52 AIDA inserted ~4cm into DTAS
60Co (~40kBq) + 137Cs (~8kBq) sources located approximately 4cm in front of the front face of the snout
10.54 DAQ start May19/R6
good event statistics OK - see attachments 11
*all* FEE64s <100k, 9x < 20k
data rate 4.9Mb/s 0.64M data items/s
11.04 nnaida17 good events c. 350k returns to 13k following Check ASIC Control (nnaida1 OK - see
https://elog.ph.ed.ac.uk/AIDA/801 )
AIDA disc statistics indicate that fast discriminators were source of higher rate - next time this
happens we should check HitRate to see whether this is a specific channel or channels and possibly
mask
11.17 DTAS DAQ issues correlation scaler reset
AIDA data file May19/R6_3
use files *from* R6_4
11.39 DAQ stop file May19/R6_8
nnrpi1 is offline - we suspect wrench or power tool placed too close to RPi when we moved AIDA
into DTAS and we jostled one or other of the cables connected to nnrpi1
nnrpi1 power cycled to reboot
*all* FEE64s OK
USB ac mains relay controls recovered OK
CAEN N1419 contrs recovered OK
11.48 AIDA withdrawn from DTAS for installation of upstream (CH2)n shield and additional Pb wall
11.52 DAQ starts to file May19/R7
alpha background resumes
ASIC settings 2018Oct03-13.26.58
*except* slow comparator 0xa -> 0x64
BNC PB-4 Pulser OFF
18.06 DAQ continues ok, on file May19/R7_0
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachment 13
FEE64 temperatures OK - see attachment 14
good event statistics OK - see attachment 15
Merger and TapeServer ok
Temperature +25.0 deg C dew point +4.4 deg C RH 27.0%
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|