Message ID: 804
Entry time: Sat May 18 04:47:22 2019
|
Author: |
TD |
Subject: |
Saturday 18 May |
|
|
12.47 DAQ continues OK to file May19/R7_3
alpha background
ASIC settings 2018Oct03-13.26.58
*except* slow comparator 0xa -> 0x64
BNC PB-4 Pulser OFF
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachments 4
Merger OK - see attachment 5
TapeServer OK - see attachment 6
12.51 Check ASIC Control *all* ASICs & FEE64s
1.8.L spectra - see attachments 7 & 8
p+n FEE64s 1.8.W spectra 20us, 200us, 2ms, 20ms FSR - attachments 9-12
n+n FEE64s 1.8.W spectra 20us, 200us, 2ms, 20ms FSR - attachments 13-16
13.13 analysis of May19/R7_3 - see attachment 17
deadtime OK
ADC data for FEE64s nnaida5 & nnaida6 hot channels (Stat spectrum channel # 104)
Also observed at a much lower rate for nnaida1 & nnaida2
13.27 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C - water topped up to H level marker
Temperature +24.6 deg C dew point +7.1 deg C RH 32.6% |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|