Message ID: 806
Entry time: Sun May 19 02:28:07 2019
|
Author: |
OH, TD, CA |
Subject: |
Sunday 19 May |
|
|
10:28 DAQ Continues ok
Alpha run continues - Currently on file R7_5
System wide checks all ok
leakage currents are up from yesterday - attachment 1
Temperature is up 2 deg C from yesterday which explains leakage current increase
Temp +26.6 deg C, Humidity 28.8% AND D.P +7.2 deg C
FEE Temperatures ok - attachment 2
Good event statistics ok - attachment 3
17.40 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C - water level 75% H-L markers
Temperature +25.4 deg C dew point +7.2 deg C RH 30.9%
17.47 DAQ continues OK to file May19/R7_6
alpha background
ASIC settings 2018Oct03-13.26.58
*except* slow comparator 0xa -> 0x64
BNC PB-4 Pulser OFF
all system wide checks OK *except*
nnaida7 fails ADC calibration
Check SYNC errors not counted since Baseline - see attachment 4 - also failed test following Collect All
SYNC Count Errors for Baseline
N.B. Check SYNC pulses received OK
detector biases & leakage currents OK - see attachments 5 & 6
FEE64 temperatures OK - see attachment 7
good event statistics OK - see attachments 8
Merger OK - see attachment 9
TapeServer OK - see attachment 10
17.53 DAQ stopped file May19/R7_6
DAQ RESET
DAQ SETUP
all system wide checks OK *except* nnaida7 fails ADC calibration
Sync all ASIC clocks using Re-Sync
Check ASIC Control *all* ASICs & FEE64s
17.57 DAQ start file May19/R8
alpha background
ASIC settings 2019May19-17.54.03
slow comparator 0x64
13.13 analysis of May19/R7_6 - see attachment 11
deadtime OK
ADC data for FEE64s nnaida5 & nnaida6 hot channels (Stat spectrum channel # 104)
Also observed at a much lower rate for nnaida1 & nnaida2 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|