AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
AIDA |
 |
|
Message ID: 828
Entry time: Tue Jun 4 07:10:27 2019
|
Author: |
TD |
Subject: |
Tuesday 4 June |
|
|
15.01 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
Water level c. 70% M-H markers - topup to 100% with 1x jugful water
Temperature +25.8 deg C dew point +8.0 deg C RH 32.4%
Access to F11 from 14.00 whilst F3 & F7 plastic scintillator and PPAC are replaced due to rad damage
during online experiment
DAQ continues OK
ASIC settings 2019May19-17.54.03
slow comparator 0xa
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachments 4
13x <20k 21x < 50k 24x <100k cf. 15x <20k, 21x < 50k, 24x < 100k yesterday evening
Merger OK - see attachment 5
TapeServer OK - see attachment 6
p+n FEE64s 1.8.W spectra 20us FSR - attachment 7
n+n FEE64s 1.8.W spectra 20us FSR - attachment 8
1.8.L spectra - see attachments 9 & 10
nnaida1 75 ch FWHM, nnaida3 82 ch FWHM - cf. 95 & 99 ch FWHM respectively yesterday afternoon
FEE64 Rate spectra - attachment 11
Check ASIC Control *all* FEE64s & ASICs x2
FEE64 Rate spectra - attachment 12 - some channels recovered
ADC data synchronised - attachment 11 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|