Message ID: 847
Entry time: Sun Jun 16 16:00:58 2019
|
Author: |
ML, CG, OH |
Subject: |
Monday 17 June 2019 |
|
|
It's a new day, it's a new log entry !!
01:25 DTAS run 46 started
BigRIPS run 2019
AIDA run R2_268
DAQ running smoothly
No change in Stat and Rate (stat zeroed at the beginning of this DTAS run)- See attachment 1
Merger rate: 630k item/sec - See Attachment 2
TapeServer rate still ~5 Mbytes/sec - See attachment 3
Temperature and HV ok - See attachment 5 & 6
System-wide checks OK except nnaida7 fails ADC calibration as usual.
!!!
Victor said that the run AIDA R2_249 is missing in the correlation merge data. This is the run when the
correlation scaler was reset (as DTAS was running out of range in some of their timing histograms) (See
elog end of Sunday)
!!!
02:23 Rate in DSSD#3 card 9 and 12 slowly reaching 100kHz while 14 and 15 ~ 50kHz. See attachment 7
TapeServer rate ~ 5.5 kHz
03:25 Change of DTAS run:
DTAS run 49 started
BigRIPS run 2020
AIDA run R2_286
Experiment still running smoothly. No significant change since last comments 2h ago.
Leakage current in detector 3 is increasing slowly see attachment 8 & 9
05:25 Change of DTAS run:
Note: DTAS run stopped during AIDA R2_305 and re-started during R2_306.
DTAS run 52 started
BigRIPS run 2021
AIDA run R2_306
Pretty quiet night shift.
Stat & Rate in attachment 10 - DSSD#3 FEE card 9 back down to 60kHz
Merger rate: 650k item/sec - See attachment 11
TapeServer rate still ~5.1 Mbytes/sec - See attachment 12
System-wide checks OK except nnaida7 fails ADC calibration as usual.
X vs Y HEC spectrum in attachment 13
LEC E vs HEC-LEC spectrum in attachment 14
Both spectra zeroed
Bias in attachment 15
Temperature unchanged
07:30 DTAS and BigRIPS stop runs
AIDA on file R2_323
07:33 New Run numbers
DTAS Run: 55
BigRIPS Run: 2022
AIDA on File: R2_324
07:56 System wide checks ok
Fee temps ok - attachment 16
100Sn setting, no degrader
ASIC settings 2019Jun13-08.56.35
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range, variable min. c. 2.5Hz
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse
08:28 Bias and leakage ok - attachment 17
08:48 ASIC check successful
09:32 DTAS and BigRIPS stop
AIDA on file R2_342
DTAS starts file 58
BigRIPS run 2023
AIDA still on file R2_342 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|