Message ID: 853
Entry time: Tue Jun 18 16:14:03 2019
|
Author: |
CA |
Subject: |
Wednesday 19 June 00.00 - 08.00 shift |
|
|
00.00 Shift Change
DAQ continues ok
AIDA file RIBF147/R2_697
DTAS run 118
BigRips run 2048
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range, variable min. c. 2.5Hz
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse
0.3mm Al degrader
00.15 System wide checks all ok *except* naida7 fails ADC calibration
00.19 FEE64 temperatures ok - attachment 1
good event statistics ok - attachment 2
detector bias/leakage currents ok - attachment 3
tape service ok - attachment 4
merger ok - attachment 5
01.19 DTAS stop/start (verified with DTAS2019 ELOG)
DTAS Stop: 01:19
DTAS run: 121
AIDA run: 708
BigRIPS run: 2049
DTAS Start: 01:19
DTAS run: 122
AIDA run: 708
BigRIPS run: 2049
01.46 Primary beam at 120 pnA; implants around 250 cps at F11
01.48 High Energy Spectra (1.8.H) for front and back strips, respectively (attachments 6 & 7)
02.35 HEC XY hit pattern - implants mostly in DSSD2 - attachment 8
HEC multiplicity n vs p - attachment 9
HEC Ey vs Ex - attachment 10
Implant - decay time vs Energy - attachment 11
02.46 zeroed all histograms, online sort stats zeroed
02.49 HEC XY hit pattern - ~3 minutes after zeroing - attachment 12
02.52 Primary beam at 120 pnA; implants around 250 cps at F11
03.20 analyser output file R2_726 - attachment 13
no timewarps
~1s deadtime on nnaida19
~0.8s deadtime on nnaida24
0s for rest
03.23 DTAS stop run 125
AIDA file R2_728
BigRIPS run 2049
DTAS start run 126
AIDA file R2_728
BigRIPS run 2050
03.28 Check ASIC control
04.16 HEC XY hit pattern - less implants than expected - attachment 14
HEC XY hit pattern - change refresh rate to 10s.... looks ok. Sorry Tom! - attachment 15
04.26 FEE64 temperatures ok - attachment 16
04.28 System wide checks all ok *except* naida7 fails ADC calibration
good event statistics ok - attachment 17
detector bias/leakage currents ok - attachment 18
merger & TapeServer ok - attachments 19 & 20
05.32 Primary beam at 120 pnA; implants around 250 cps at F11
05.36 DTAS stop run 129
AIDA file R2_749
BigRIPS run 2050
05.40 DTAS send sync pulse AIDA file R2_750
05.43 DTAS start run 130
AIDA file R2_751
BigRIPS run 2051
06.31 HEC XY hit pattern - implants mostly in DSSD2 - attachment 21
HEC multiplicity n vs p - attachment 22
HEC Ey vs Ex - attachment 23
Implant - decay time vs Energy - attachment 24
06.34 High Energy Spectra (1.8.H) for front and back strips, respectively (attachments 25 & 26)
06.36 3.20 analyser output file R2_731 - attachment 27
no timewarps
~1.7s deadtime on nnaida19
~0.8s deadtime on nnaida24
0s for rest
07.19 System wide checks all ok *except* naida7 fails ADC calibration
good event statistics ok - attachment 28
FEE64 temperatures ok - attachment 29
detector bias/leakage currents ok - attachment 30
07.21 Merger & TapeServer ok - attachments 31 & 32
07.30 DTAS stop run 132
AIDA file R2_768
BigRIPS run 2051
DTAS start run 133
AIDA file R2_768
BigRIPS run 2052
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|