AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  AIDA  ELOG logo
Message ID: 862     Entry time: Tue Feb 9 11:44:21 2021
Author: OH, TD, CA, LS 
Subject: 9th Feb GSI Dry run 
Database file - 2019Dec19-16.19.51
Current threshold at 0x64


13:27 Recovered from power cycle
      System wide checks all ok - Attachments 1-6
      *NB* Synchronise the ASIC clocks doesn't work - Attachment 7
      ASIC Control ok
      FEE temperatures ok - Attachment 8

      Bias and leakage currents ok - Attachment 9
      New taper service menu - Attachment 10

      Pulser peaks log plot - Attachments 11-12
      1.8.L pulser peaks - Attachment 13-14
      Waveform plots - Attachment 15-16



      FEE7 we aren't seeing a HIT pattern but we are seeing ADC data items -> It was a corrupted options file

15:20 DAQ Crashed after enabling data transfer
      Problem was not configuring the tapeserver correctly
      Have recovered from this. There are now no histograms for FEE09
      This is because the options file has been corrupted during startup.
      From the statistics it appear we are still reading data on the FEE
      Chosen to let the DAQ run in this state to test stability for the time being
Attachment 1: SystemWideChecks1145.PNG  179 kB  Uploaded Tue Feb 9 15:40:18 2021  | Hide | Hide all
SystemWideChecks1145.PNG
Attachment 2: ClockStatus1145_1.PNG  236 kB  Uploaded Tue Feb 9 15:42:08 2021  | Hide | Hide all
ClockStatus1145_1.PNG
Attachment 3: ADC_Calibration_1146.PNG  214 kB  Uploaded Tue Feb 9 15:42:18 2021  | Hide | Hide all
ADC_Calibration_1146.PNG
Attachment 4: SystemWideChecks1147.PNG  229 kB  Uploaded Tue Feb 9 15:42:45 2021  | Hide | Hide all
SystemWideChecks1147.PNG
Attachment 5: 1148_SystemWideChecksFPGA.PNG  218 kB  Uploaded Tue Feb 9 15:42:57 2021  | Hide | Hide all
1148_SystemWideChecksFPGA.PNG
Attachment 6: 1148_MemInfo.PNG  254 kB  Uploaded Tue Feb 9 15:43:02 2021  | Hide | Hide all
1148_MemInfo.PNG
Attachment 7: 1152_SychroniseASICClocks.PNG  253 kB  Uploaded Tue Feb 9 15:43:16 2021  | Hide | Hide all
1152_SychroniseASICClocks.PNG
Attachment 8: 1153_FeeTemp.PNG  303 kB  Uploaded Tue Feb 9 15:43:16 2021  | Hide | Hide all
1153_FeeTemp.PNG
Attachment 9: 1229_Bias.PNG  55 kB  Uploaded Tue Feb 9 15:43:16 2021  | Hide | Hide all
1229_Bias.PNG
Attachment 10: 1233_TapeService.PNG  315 kB  Uploaded Tue Feb 9 15:43:16 2021  | Hide | Hide all
1233_TapeService.PNG
Attachment 11: 1320_1_8_L_Odd.PNG  358 kB  Uploaded Tue Feb 9 15:43:16 2021  | Hide | Hide all
1320_1_8_L_Odd.PNG
Attachment 12: 1322_1_8_L_Even.PNG  374 kB  Uploaded Tue Feb 9 15:43:16 2021  | Hide | Hide all
1322_1_8_L_Even.PNG
Attachment 13: 1326_1_8_L_Odd_05_100_Width.PNG  367 kB  Uploaded Tue Feb 9 15:43:16 2021  | Hide | Hide all
1326_1_8_L_Odd_05_100_Width.PNG
Attachment 14: 1327_1_8_L_Even_08_90.PNG  396 kB  Uploaded Tue Feb 9 15:43:16 2021  | Hide | Hide all
1327_1_8_L_Even_08_90.PNG
Attachment 15: 1332_Waveform_odd.PNG  607 kB  Uploaded Tue Feb 9 15:43:16 2021  | Hide | Hide all
1332_Waveform_odd.PNG
Attachment 16: 1333_Waveform_Even.PNG  572 kB  Uploaded Tue Feb 9 15:43:16 2021  | Hide | Hide all
1333_Waveform_Even.PNG
ELOG V3.1.4-unknown