Test of AIDA FEE64 d8:80:39:41:f6:ee ( aida01 ) - HDMI connector replaced |
|
|
Test of AIDA FEE64 d8:80:39:41:f6:ee ( aida01 ) - HDMI connector replaced. Cooling plate fitted - no base plate.
ISOL test station STFC DL, T9
Pulser BNC PB-5
Amplitude 0.5V
Attenuation x1
Polarity +
Frequency 100Hz
Tail pulse, decay time 1ms
Delay 250ns (min)
ASIC settings 2019May22-10.19.28 - attachment 1
Slow comparator 0x80
Options - attachment 2
Adaptor PCB - PJCS
FEE64 temperatures OK - attachments 3
System wide checks - attachments 4-9
FEE64 module aida01 global clocks failed, 6
Clock status test result: Passed 3, Failed 1
FEE64 module aida01 failed
Calibration test result: Passed 3, Failed 1
Timestamp RAM values OK?
Statistics OK - attachments 10-11
Good WAVE events reported?
LEC spectra - attachments 12-15
Data observed for all channels of all ASICs - double peaking observed - may be due to incomplete FEE64 assembly? See attachment 17
Hit/HitRate spectra - attachment 16
OK *except*
1.2.L fast comparator missing
no WAVE data |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|